ArticleEuropean spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society2015
Construct stability of an instrumented 2-level cervical corpectomy model following fatigue testing: biomechanical comparison of circumferential antero-posterior instrumentation versus a novel anterior-only transpedicular screw-plate fixation technique.
Article in European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society, 2015. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 16 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
16 citing papers in PubMed, 34 citations in OpenAlex.
- Circumferential Surgical Management of a Cervical Chordoma: A Case Report and Review of the Literature.Cureus · 2025Article
- Radiological studies for the best position and trajectory of the anterior cervical pedicle screw in the lower cervical spine in normal populations.Quantitative imaging in medicine and surgery · 2025Article
- Pars interarticularis screws for posterior cervical fusion - investigating a new trajectory using a CT-based multiplanar reconstruction: Part I.Acta neurochirurgica · 2024Article
- Biomechanical study of anterior transpedicular root screw intervertebral fusion system of lower cervical spine: a finite element analysis.Frontiers in bioengineering and biotechnology · 2024Article
- Measurement of anatomical parameters of anterior transpedicular root screw intervertebral fusion system of cervical spine.BMC musculoskeletal disorders · 2023Article
- Comparison of anterior column reconstruction techniques after en bloc spondylectomy: a finite element study.Scientific reports · 2023Article
- Clinical and surgical results related to anterior-only multilevel cervical decompression and instrumented fusion for degenerative disease.Brain & spine · 2023Article
- Circumferential Operations of the Cervical Spine.Neurospine · 2021Article
- Article
- Dual pitch titanium-coated pedicle screws improve initial and early fixation in a polyetheretherketone rod semi-rigid fixation system in sheep.Chinese medical journal · 2019Article
- Progress of the Anterior Transpedicular Screw in Lower Cervical Spine: A Review.Medical science monitor : international medical journal of experimental and clinical research · 2019Review
- Two-level cervical corpectomy-long-term follow-up reveals the high rate of material failure in patients, who received an anterior approach only.Neurosurgical review · 2019Article
- Retrospective analysis of cervical corpectomies: implant-related complications of one- and two-level corpectomies in 45 patients.Neurosurgical review · 2018Article
- Accuracy of 3D fluoro-navigated anterior transpedicular screws in the subaxial cervical spine: an experimental study on human specimens.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2017Article
- Cement-augmented screws in a cervical two-level corpectomy with anterior titanium mesh cage reconstruction: a biomechanical study.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2017Article
- Biomechanical testing of circumferential instrumentation after cervical multilevel corpectomy.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2015Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
10 authors at 3 institutions in 2 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
introductionA high rate of complications in multilevel cervical surgery with corpectomies and anterior-only screw-and-plate stabilization is reported. A 360°-instrumentation improves construct stiffness and fusion rates, but adds the morbidity of a second approach. A novel ATS-technique (technique that used anterior transpedicular screw placement) was recently described, yet no study to date has analyzed its performance after fatigue loading. Accordingly, the authors performed an analysis of construct stiffness after fatigue testing of a cervical 2-level corpectomy model reconstructed using a novel anterior transpedicular screw-and-plate technique (ATS-group) in comparison to standard antero-posterior instrumentation (360°-group). MATERIALS AND
methodsTwelve fresh-frozen human cervical spines were mounted on a spine motion tester to analyze restriction of ROM under loading (1.5 Nm) in flexion-extension (FE), axial rotation (AR), and lateral bending (LB). Testing was performed in the intact state, and after instrumentation of a 2-level corpectomy C4 + C5 using a cage and the constructs of ATS- and 360°-group, after 1,000 cycles, and after 2,000 cycles of fatigue testing. In the ATS-group (n = 6), instrumentation was achieved using a customized C3-C6 ATS-plate system. In the 360°-group (n = 6), instrumentation consisted of a standard anterior screw-and-plate system with a posterior instrumentation using C3-C6 lateral mass screws. Motion data were assessed as degrees and further processed as normalized values after standardization to the intact ROM state.
resultsSpecimen age and BMD were not significantly different between the ATS- and 360°-groups. After instrumentation and 2,000 cycles of testing, no specimen exhibited a ROM greater than in the intact state. No specimen exhibited catastrophic construct failure after 2,000 cycles. Construct stiffness in the 360°-group was significantly increased compared to the ATS-group for all loading conditions, except for FE-testing after instrumentation. After 2,000 cycles, restriction of ROM under loading in FE was 39.8 ± 30% in the ATS-group vs. 2.8 ± 2.3% in the 360°-group, in AR 60.4 ± 25.8 vs 15 ± 11%, and in LB 40 ± 23.4 vs 3.9 ± 1.2%. Differences were significant (p < 0.05).
conclusion360°-instrumentation resembles the biomechanical standard of reference for stabilization of 2-level corpectomies. An ATS-construct was also shown to confer high construct stiffness, significantly reducing the percentage ROM beyond that of an intact specimen after 2,000 cycles. This type of instrumentation might be a clinical valuable and biomechanically sound adjunct to multilevel anterior surgical procedures.