Evidence map›Paper›PMID 25520417›Full record

ArticleThe Journal of bone and joint surgery. American volume2014

Comparative effectiveness of ceramic-on-ceramic implants in stemmed hip replacement: a multinational study of six national and regional registries.

Art Sedrakyan, Stephen Graves, Barbara Bordini, Miquel Pons, Leif Havelin, Susan Mehle, Elizabeth Paxton, Thomas Barber, Guy Cafri

Abstract readComparative Study
In one paragraph

Article in The Journal of bone and joint surgery. American volume, 2014. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.

0numbers the graph read from it
0cells of the map it votes in
8citing papers in PubMed
2.9field-weighted citation impact, top 9% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

8 citing papers in PubMed, 26 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Are powder-technology-built stems safe? A midterm follow-up registry study.Journal of materials science. Materials in medicine · 2021
    Article
  5. Article
  6. Article
  7. Article
  8. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors at 8 institutions in 5 countries.

Art SedrakyanWeill Cornell Medical College, 402 East 67th Street, New York, NY 10065.
Stephen GravesAustralian Orthopaedic Association National Joint Replacement Registry, Discipline of Public Health, MDP DX 650 511, University of Adelaide, Adelaide, SA 5005, Australia.
Barbara BordiniRegister of Orthopaedic Implants (RIPO), c/o Medical Technology Laboratory, Istituto Ortopedico Rizzoli, via de Barbiano 1/10, 40136 Bologna, Italy.
Miquel PonsHip and Knee Replacement Unit, Orthopaedic Surgery Department, Hospital Sant Rafael, Paseo Vall d'Hebrón 107-117, Barcelona 08035, Spain.
Leif HavelinThe Norwegian Arthroplasty Register, Department of Orthopaedic Surgery, Haukeland University Hospital, Mollendalsbakken 11, N-5021 Bergen, Norway.
Susan MehleHealthEast Joint Registry (HEJR), 1690 University Avenue West, Data Science, Suite 400, Minneapolis, MN 55104.
Elizabeth PaxtonSurgical Outcomes & Analysis Department, Kaiser Permanente, 8954 Rio San Diego Drive, Suite 406, San Diego, CA 92108.
Thomas BarberDepartment of Orthopedic Surgery, Kaiser Permanente, 280 West MacArthur Boulevard, Oakland, CA 94611.
Guy CafriSurgical Outcomes & Analysis Department, Kaiser Permanente, 8954 Rio San Diego Drive, Suite 406, San Diego, CA 92108.
Kaiser Permanente San Diego Medical Center · USCornell University · USHaukeland University Hospital · NOHealthEast Care System · USHospital Universitario San Rafael · ESIstituto Ortopedico Rizzoli · ITKaiser Permanente · USUniversity of Adelaide · AU

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundThe rapid decline in use of conventional total hip replacement with a large femoral head size and a metal-on-metal bearing surface might lead to increased popularity of ceramic-on-ceramic bearings as another hard-on-hard alternative that allows implantation of a larger head. We sought to address comparative effectiveness of ceramic-on-ceramic and metal-on-HXLPE (highly cross-linked polyethylene) implants by utilizing the distributed health data network of the ICOR (International Consortium of Orthopaedic Registries), an unprecedented collaboration of national and regional registries and the U.S. FDA (Food and Drug Administration).

methodsA distributed health data network was developed by the ICOR and used in this study. The data from each registry are standardized and provided at a level of aggregation most suitable for the detailed analysis of interest. The data are combined across registries for comprehensive assessments. The ICOR coordinating center and study steering committee defined the inclusion criteria for this study as total hip arthroplasty performed without cement from 2001 to 2010 in patients forty-five to sixty-four years of age with osteoarthritis. Six national and regional registries (Kaiser Permanente and HealthEast in the U.S., Emilia-Romagna region in Italy, Catalan region in Spain, Norway, and Australia) participated in this study. Multivariate meta-analysis was performed with use of linear mixed models, with survival probability as the unit of analysis. We present the results of the fixed-effects model and include the results of the random-effects model in an appendix. SAS version 9.2 was used for all analyses. We first compared femoral head sizes of >28 mm and ≤28 mm within ceramic-on-ceramic implants and then compared ceramic-on-ceramic with metal-on-HXLPE.

resultsA total of 34,985 patients were included; 52% were female. We found a lower risk of revision associated with use of ceramic-on-ceramic implants when a larger head size was used (HR [hazard ratio] = 0.73, 95% CI [confidence interval] = 0.60 to 0.88, p = 0.001). Use of smaller-head-size ceramic-on-ceramic bearings was associated with a higher risk of failure compared with metal-on-HXLPE bearings (HR = 1.36, 95% CI = 1.09 to 1.68, p = 0.006). Use of large-head-size ceramic-on-ceramic bearings was associated with a small protective effect relative to metal-on-HXLPE bearings (not subdivided by head size) in years zero to two, but this difference dissipated over the longer term.

conclusionsOur multinational study based on a harmonized, distributed network showed that use of ceramic-on-ceramic implants with a smaller head size in total hip arthroplasty without cement was associated with a higher risk of revision compared with metal-on-HXLPE and >28-mm ceramic-on-ceramic implants. These findings warrant careful reflection by regulatory and clinical communities and wide dissemination to patients for informed decision-making regarding such surgery.

Indexed as

Hip ProsthesisProsthesis DesignArthroplasty, Replacement, HipCeramicsFemaleFemur HeadHumansMaleMiddle AgedOsteoarthritis, HipProsthesis FailureRegistriesReoperation

Identifiers

PMID25520417
PMCPMC4271430
OpenAlexW2154242740

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked