Evidence map›Paper›PMID 25337130›Full record

ArticleJournal of contemporary brachytherapy2014

The impact of activating source dwell positions outside the CTV on the dose to treated normal tissue volumes in TRUS guided 3D conformal interstitial HDR brachytherapy of prostate cancer.

Leif Karlsson, Per Thunberg, Bengt Johansson, Jan Persliden

Abstract read
In one paragraph

Article in Journal of contemporary brachytherapy, 2014. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.

0numbers the graph read from it
0cells of the map it votes in
5citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

5 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Leif KarlssonDepartment of Medical Physics, Örebro University Hospital, Örebro, Sweden ; School of Health and Medical Sciences, Örebro University, Örebro, Sweden.
Per ThunbergDepartment of Medical Physics, Örebro University Hospital, Örebro, Sweden ; School of Health and Medical Sciences, Örebro University, Örebro, Sweden.
Bengt JohanssonDepartment of Oncology, Örebro University Hospital, Örebro, Sweden.
Jan PerslidenDepartment of Medical Physics, Örebro University Hospital, Örebro, Sweden ; School of Health and Medical Sciences, Örebro University, Örebro, Sweden.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeDose coverage is crucial for successful treatment in mono-brachytherapy. Since few and very high dose fractions are used, there is an important balance between dwell positioning outside the clinical target volume (CTV) and possible damage on adjacent normal tissue. The purpose of this study was to evaluate the possibility of having dwell positions close to the CTV surface, while maintaining an acceptable dose distribution, and to investigate the robustness in terms of known geometrical uncertainties of the implant. MATERIAL AND

methodsThis study included 37 patients who had received brachytherapy for prostate cancer as a monotherapy with the following schedules: 2 × 14 Gy or 3 × 11 Gy, each fraction separated by two weeks. The source dwell positions were activated 5 mm outside CTV. New optimizations were simulated for dwell positions at 3, 2, 1, and 0 mm. Inverse and graphical optimization were applied according to the relative dose constraints: V100 CTV ≥ 97%, Dmax, urethra ≤ 110%, and D10 rectal mucosa ≤ 65%. The V100, normal tissue outside CTV was used to evaluate dose variations caused by different dwell positions. Prostate geometries and dose distributions for the different dwell positions outside the CTV were used to investigate the impact on the CTV dose distribution due to geometrical uncertainties.

resultsBoth V100, CTV, and V100, normal tissue decreased, 98.6% to 92.2%, and 17 cm(3) to 9.0 cm(3), for dwell activation from 5 mm to 0 mm. The evaluation of both simulated longitudinal geometrical uncertainties and different source dwell activations implied that V100, CTV ranged from 98.6% to 86.3%.

conclusionsIt is possible to reduce the V100, normal tissue by decreasing the source dwell positions outside the CTV from 5 to 3 mm, while maintaining dose constraints. In combination with the estimated geometrical uncertainties, however, the source dwell positions need to be 5 mm from the surface in order to maintain a robust implant.

Indexed as

brachytherapymarginsoptimizationprostate cancer

Identifiers

PMID25337130
PMCPMC4200186