Evidence map›Paper›PMID 25270081›Full record

ArticleDer Orthopade2014

[Damage to implants due to high-frequency electrocautery : analysis of four fractured hip endoprostheses shafts].

C Konrads, M N Wente, W Plitz, M Rudert, M Hoberg

Abstract readCase ReportsEnglish Abstract
PubMed Publisher
In one paragraph

Article in Der Orthopade, 2014. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed, 1 pooled it
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed, 1 synthesis or guideline pooled it.

  1. Pooled it
  2. Article
  3. Metallic Implants Used in Lumbar Interbody Fusion.Materials (Basel, Switzerland) · 2022
    Review
  4. Article
  5. Review
  6. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

C KonradsOrthopädische Klinik König-Ludwig-Haus, Universität Würzburg, Brettreichstraße 11, 97074, Würzburg, Deutschland, christian.konrads@gmail.com.
M N Wente
W Plitz
M Rudert
M Hoberg

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

In revision surgery of joints, high-frequency electrocauterization instruments are used for homeostasis and dissection of soft tissue. If there is contact of these instruments with the metal implants, flashover can occur. This can lead to thermal microstructural changes in the material and as a consequence may reduce the fatigue strength of the implant. Four cases of hip revision surgeries were analysed. In all cases flashovers occurred and secondarily, the titanium hip endoprosthesis stem broke in the neck section of the prosthesis. The conducted investigations showed that contact between the high-frequency instrument and the anterolateral aspect of the endoprosthesis neck had occurred. Electrothermal implant damage was found in the broken area. If in hip revision surgery the stem is not to be replaced, contact between high-frequency instruments and the metal implant should be avoided.

Indexed as

ElectrocoagulationHip ProsthesisMetalsProsthesis FailureHip FracturesHumansMetals