Evidence map›Paper›PMID 25208953›Full record

SynthesisBMJ (Clinical research ed.)2014

Appraisal of evidence base for introduction of new implants in hip and knee replacement: a systematic review of five widely used device technologies.

Marc J Nieuwenhuijse, R G H H Nelissen, J W Schoones, A Sedrakyan

Open access · hybridAbstract readSystematic Review
In one paragraph

Synthesis in BMJ (Clinical research ed.), 2014. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 31 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
31citing papers in PubMed, 1 pooled it
12.3field-weighted citation impact, top 1% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

31 citing papers in PubMed, 1 synthesis or guideline pooled it, 85 citations in OpenAlex.

  1. Pooled it
  2. Trial
  3. Review
  4. Practice Patterns of Adult Reconstruction Fellowship-Trained Surgeons: Current Trends and Evolution of Training From 1986 to 2022.Journal of the American Academy of Orthopaedic Surgeons. Global research & reviews · 2025
    Article
  5. Article
  6. Review
  7. Review
  8. Article
  9. Article
  10. Article
  11. Review
  12. Article
  13. Observational
  14. Article
  15. Article
  16. Review
  17. Additive Manufacturing in Total Joint Arthroplasty.The Orthopedic clinics of North America · 2019
    Review
  18. Article
  19. Article
  20. Article
4 · The record

Corrections and comments

5 · Who and what money

Authors and funding

4 authors at 2 institutions in 2 countries.

Marc J NieuwenhuijsePatient Centered Comparative Effectiveness Program and US Food and Drug Administration Medical Device Epidemiology Science and Infrastructure Center, Department of Public Health, Weill Cornell Medical College, New York, NY 10065, USA Department of Orthopaedics, Leiden University Medical Center, 2300 RC Leiden, Netherlands Office of Surveillance and Biometrics, Center for Devices and Radiological Health, FDA, Silver Spring, MD 20993, USA.
R G H H NelissenDepartment of Orthopaedics, Leiden University Medical Center, 2300 RC Leiden, Netherlands.
J W SchoonesWalaeus Library, Leiden University Medical Center, 2300 RC Leiden, Netherlands.
A SedrakyanPatient Centered Comparative Effectiveness Program and US Food and Drug Administration Medical Device Epidemiology Science and Infrastructure Center, Department of Public Health, Weill Cornell Medical College, New York, NY 10065, USA Office of Surveillance and Biometrics, Center for Devices and Radiological Health, FDA, Silver Spring, MD 20993, USA ars2013@med.cornell.edu.
Center for Devices and Radiological Health · USLeiden University Medical Center · NL

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectiveTo determine the evidence of effectiveness and safety for introduction of five recent and ostensibly high value implantable devices in major joint replacement to illustrate the need for change and inform guidance on evidence based introduction of new implants into healthcare.

designSystematic review of clinical trials, comparative observational studies, and registries for comparative effectiveness and safety of five implantable device innovations. DATA SOURCES: PubMed (Medline), Embase, Web of Science, Cochrane, CINAHL, reference lists of articles, annual reports of major registries, summaries of safety and effectiveness for pre-market application and mandated post-market studies at the US Food and Drug Administration. STUDY SELECTION: The five selected innovations comprised three in total hip replacement (ceramic-on-ceramic bearings, modular femoral necks, and uncemented monoblock cups) and two in total knee replacement (high flexion knee replacement and gender specific knee replacement). All clinical studies of primary total hip or knee replacement for symptomatic osteoarthritis in adults that compared at least one of the clinical outcomes of interest (patient centred outcomes or complications, or both) in the new implant group and control implant group were considered. Data searching, abstraction, and analysis were independently performed and confirmed by at least two authors. Quantitative data syntheses were performed when feasible.

resultsAfter assessment of 10,557 search hits, 118 studies (94 unique study cohorts) met the inclusion criteria and reported data related to 15,384 implants in 13,164 patients. Comparative evidence per device innovation varied from four low to moderate quality retrospective studies (modular femoral necks) to 56 studies of varying quality including seven high quality (randomised) studies (high flexion knee replacement). None of the five device innovations was found to improve functional or patient reported outcomes. National registries reported two to 12 year follow-up for revision occurrence related to more than 200,000 of these implants. Reported comparative data with well established alternative devices (over 1,200,000 implants) did not show improved device survival. Moreover, we found higher revision occurrence associated with modular femoral necks (hazard ratio 1.9) and ceramic-on-ceramic bearings (hazard ratio 1.0-1.6) in hip replacement and with high flexion knee implants (hazard ratio 1.0-1.8).

conclusionWe did not find convincing high quality evidence supporting the use of five substantial, well known, and already implemented device innovations in orthopaedics. Moreover, existing devices may be safer to use in total hip or knee replacement. Improved regulation and professional society oversight are necessary to prevent patients from being further exposed to these and future innovations introduced without proper evidence of improved clinical efficacy and safety.

Indexed as

Arthroplasty, Replacement, HipArthroplasty, Replacement, KneeEpidemiologic MethodsFemaleHip ProsthesisHumansInventionsKnee ProsthesisMalePatient SafetyProsthesis DesignProsthesis FailureRange of Motion, ArticularReoperationSex Distribution

Identifiers

PMID25208953
PMCPMC4159610
OpenAlexW2002116435

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.