Evidence map›Paper›PMID 25060175›Full record

ArticleStrahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al]2014

New possibilities for volumetric-modulated arc therapy using the Agility™ 160-leaf multileaf collimator.

Nadine Blümer, Christian Scherf, Janett Köhn, Eugen Kara, Britta Loutfi-Krauß, Detlef Imhoff, Claus Rödel, Ulla Ramm, Jörg Licher

Abstract read
PubMed Publisher
In one paragraph

Article in Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al], 2014. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.

0numbers the graph read from it
0cells of the map it votes in
7citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

7 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors.

Nadine BlümerKlinik für Strahlentherapie und Onkologie, Universitätsklinikum Frankfurt, Theodor-Stern-Kai 7, 60590, Frankfurt am Main, Germany, Nadine.Bluemer@kgu.de.
Christian Scherf
Janett Köhn
Eugen Kara
Britta Loutfi-Krauß
Detlef Imhoff
Claus Rödel
Ulla Ramm
Jörg Licher

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeThis study compares the quality of intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT) plans optimized for an Elekta Agility(TM) (Elekta, Stockholm, Sweden) multileaf collimator (MLC; leaf width 5 mm) and an Elekta MLCi2 (leaf width 10 mm) for complex target volumes (anal, AC; head and neck, H&N and prostate cancer, PC). PATIENTS AND

methodsFor plan comparisons, 15 patients who had been treated with IMRT or VMAT using the MLCi2 were selected. For each patient, a retrospective treatment plan using the MLCi2 for the technique not applied was created, as were treatment plans for both techniques using the Agility(TM) MLC. Dose-volume histograms (DHVs) for planning target volumes (PTVs) and organs at risk (OARs) were compared. Further parameters relating to dose conformity, dose homogeneity and mean dose (Dmean) to the PTV, compliance with the intended OAR dose criteria and overall dose to normal tissue were analyzed. Verification measurements were performed and optimization and treatment times were compared.

resultsCompared to the MLCi2 plans, the Agility(TM) IMRT and VMAT plans show better or equivalent results in terms of PTV dose conformity and homogeneity. Compliance with the intended OAR dose criteria does not differ according to technique or MLC type. Slight differences are shown for dose distributions in OARs and normal tissue. Verification measurements show that all plans fulfill the acceptance criteria of a minimum of 95 % matched dose points for the 3 %/3 mm γ criterion. Optimization times for the VMAT plans increase compared to the IMRT plans, whereas treatment times decrease.

conclusionWith the MLCi2, treatment of complex target volumes with VMAT was only possible with compromises in terms of target coverage. Using the Agility(TM) MLC, even complex target volumes can be treated with VMAT without compromising target coverage or resulting in higher exposure of OARs or normal tissue.

Indexed as

Equipment DesignEquipment Failure AnalysisHumansNeoplasmsRadiotherapy DosageRadiotherapy, Intensity-ModulatedRadiotherapy Planning, Computer-AssistedReproducibility of ResultsRetrospective StudiesSensitivity and SpecificityTreatment Outcome

Identifiers

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked