ISO-based assessment of accuracy and precision of glucose meters in dogs.
Article in Journal of veterinary internal medicine. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
9 citing papers in PubMed.
- Analytical Evaluation of Three Portable Blood Glucose Meters for Human Use in Dogs.Veterinary sciences · 2025Article
- Impact of haematocrit on the accuracy of spot blood glucose measurements in dogs using two veterinary hand-held glucometers.Frontiers in veterinary science · 2025Article
- High carbohydrate is preferable to high lipid parenteral nutrition in healthy dogs undergoing prolonged sedation.Veterinary research communications · 2024Article
- Impact on result interpretation of correct and incorrect selection of veterinary glucometer canine and feline settings.Journal of veterinary diagnostic investigation : official publication of the American Association of Veterinary Laboratory Diagnosticians, Inc · 2023Article
- Evaluation of Three Human-Use Glucometers for Blood Glucose Measurement in Dogs.Veterinary medicine international · 2022Article
- Comparison of glucose concentrations in canine whole blood, plasma, and serum measured with a veterinary point-of-care glucometer.Journal of veterinary diagnostic investigation : official publication of the American Association of Veterinary Laboratory Diagnosticians, Inc · 2021Article
- Efficacy and Safety of Intramuscular Insulin Lispro vs. Continuous Intravenous Regular Insulin for the Treatment of Dogs With Diabetic Ketoacidosis.Frontiers in veterinary science · 2020Article
- Evaluation of newly developed veterinary portable blood glucose meter with hematocrit correction in dogs and cats.The Journal of veterinary medical science · 2017Article
- Accuracy of a Flash Glucose Monitoring System in Diabetic Dogs.Journal of veterinary internal medicine · 2016Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundPortable blood glucose meters (PBGMs) allow easy glucose measurements. As animal-specific PBGMs are not available everywhere, those for humans are widely used.
objectivesTo assess the accuracy and precision of 9 PBGMs in canine whole blood (WB) and plasma, based on the ISO 15197:2013. ANIMALS: Fifty-nine client-owned dogs attending the Veterinary Teaching Hospital.
methodsAnalytical evaluation of 100 blood samples was performed for accuracy and 23 for precision (glucose 29-579 mg/dL) following ISO recommendations. A PBGM was considered accurate if 95% of the measurements were within ±15 mg/dL from the reference when glucose was <100 mg/dL and within ±15% when it was ≥100 mg/dL, and if 99% of them were within zones A and B in error grid analysis (EG). A hexokinase-based analyzer was used as reference. Ninety samples were assessed for hematocrit interferences.
resultsAccuracy requirements were not fulfilled by any PBGM in WB (74% of measurements within the limits for the most accurate) and by 1 only in plasma. However, the EG analysis in WB was passed by 6 PBGM and by all in plasma. The most accurate were also the most precise, with coefficients of variation <5% in WB and <3% in plasma. Hematocrit correlated with bias against the reference method in 4 PBGM (r = -0.243 - [-0.371]; P < .021). CONCLUSIONS AND CLINICAL IMPORTANCE: This disparity among PBGM suggests that meters approved for humans need to be evaluated before use in other species.