Evidence map›Paper›PMID 24854520›Full record

ArticleJournal of anesthesia2014

Condensation of humidified air in the inflation line of a polyurethane cuff precludes correct continuous pressure monitoring during mechanical ventilation.

Herbert Spapen, Walter Moeyersons, Wim Stiers, Geert Desmet, Emiel Suys

Abstract read
PubMed Publisher
In one paragraph

Article in Journal of anesthesia, 2014. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Article
  2. Review
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Herbert SpapenIntensive Care Department, University Hospital, Vrije Universiteit Brussels, Laarbeeklaan, 101, 1090, Brussels, Belgium, h.spapen@skynet.be.
Walter Moeyersons
Wim Stiers
Geert Desmet
Emiel Suys

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Within continuously controlled limits of cuff pressure, an endotracheal tube cuff made of polyurethane (PU) prevents secretion inflow better than a conventional polyvinylchloride cuff. However, the temperature difference between ventilator gas and the air inside a PU cuff produces condensation droplets that accumulate in the cuff inflation canal. We investigated whether condensation influenced continuous cuff pressure monitoring. A PU-cuffed tube was placed into an artificial trachea and connected to a ventilator and test lung. An additional line was inserted at the distal cuff end to directly measure pressure inside the cuff. Methylene blue instillation via the inflation line mimicked condensation. Inspiratory (Pinsp) and expiratory (Pexp) pressures were continuously recorded. Six consecutive experiments were performed comparing pressures at baseline (T0) and at 24 h (T24). Shortly after dye instillation, pressures recorded at the inflation canal became permanently fixed at 25 cmH2O. In contrast, pressures measured directly in the cuff progressively decreased (mean Pinsp 30 ± 3 vs. 20 ± 2 cmH2O and mean Pexp 25 ± 0 vs. 12 ± 2 cmH2O, T0 vs. T24, both P < 0.05). Thus, condensation in the inflation line of a PU-cuffed tube renders continuous cuff pressure monitoring unreliable, thereby increasing the risk for microaspiration.

Indexed as

Intubation, IntratrachealPolyurethanesPressureRespiration, ArtificialTracheaVentilators, MechanicalPolyurethanes