Evidence map›Paper›PMID 24847790›Full record

ArticleActa orthopaedica2014

No association between serum metal ions and implant fixation in large-head metal-on-metal total hip arthroplasty.

Mette Holm Hjorth, Kjeld Søballe, Stig Storgaard Jakobsen, Nina Dyrberg Lorenzen, Inger Mechlenburg, Maiken Stilling

Open access · goldAbstract read
In one paragraph

Article in Acta orthopaedica, 2014. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
7citing papers in PubMed, 1 pooled it
1.2field-weighted citation impact, top 22% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

7 citing papers in PubMed, 1 synthesis or guideline pooled it, 10 citations in OpenAlex.

  1. Pooled it
  2. Trial
  3. Article
  4. Article
  5. Antioxidant System and Ions Concentrations After Femoral Neck and Resurfacing Hip Arthroplasty.Medical science monitor : international medical journal of experimental and clinical research · 2022
    Article
  6. Article
  7. [Mid- and long-term effectiveness and failure causes analysis of large-head metal-on-metal total hip arthroplasty].Zhongguo xiu fu chong jian wai ke za zhi = Zhongguo xiufu chongjian waike zazhi = Chinese journal of reparative and reconstructive surgery · 2017
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 1 institution in 1 country.

Mette Holm HjorthDepartment of Orthopedic Surgery , Aarhus University Hospital, Aarhus , Denmark.
Kjeld Søballe
Stig Storgaard Jakobsen
Nina Dyrberg Lorenzen
Inger Mechlenburg
Maiken Stilling
Aarhus University Hospital · DK

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundThe mechanism of failure of metal-on-metal (MoM) total hip arthroplasty (THA) has been related to a high rate of metal wear debris, which is partly generated from the head-trunnion interface. However, it is not known whether implant fixation is affected by metal wear debris. PATIENTS AND

methods49 cases of MoM THA in 41 patients (10 women) with a mean age of 52 (28-68) years were followed with stereoradiographs after surgery and at 1, 2, and 5 years to analyze implant migration by radiostereometric analysis (RSA). Patients also participated in a 5- to 7-year follow-up with measurement of serum metal ions, questionnaires (Oxford hip score (OHS) and Harris hip score (HHS)), and measurement of cup and stem positions and systemic bone mineral density.

resultsAt 1-2 years, mean total translation (TT) was 0.04 mm (95% CI: -0.07 to 0.14; p = 0.5) for the stems; at 2-5 years, mean TT was 0.13 mm (95% CI: -0.25 to -0.01; p = 0.03), but within the precision limit of the method. For the cups, there was no statistically significant TT or total rotation (TR) at 1-2 and 2-5 years. At 2-5 years, we found 4 cups and 5 stems with TT migrations exceeding the precision limit of the method. There was an association between cup migration and total OHS < 40 (4 patients, 4 hips; p = 0.04), but there were no statistically significant associations between cup or stem migration and T-scores < -1 (n = 10), cup and stem positions, or elevated serum metal ion levels (> 7µg/L (4 patients, 6 hips)).

interpretationMost cups and stems were well-fixed at 1-5 years. However, at 2-5 years, 4 cups and 5 stems had TT migrations above the precision limits, but these patients had serum metal ion levels similar to those of patients without measurable migrations, and they were pain-free. Patients with serum metal ion levels > 7 µg/L had migrations similar to those in patients with serum metal ion levels < 7 µg/L. Metal wear debris does not appear to influence the fixation of hip components in large-head MoM articulations at medium-term follow-up.

Indexed as

Absorptiometry, PhotonAcetabulumAdultAgedArthroplasty, Replacement, HipFemaleFollow-Up StudiesForeign-Body MigrationHip JointHip ProsthesisHumansIonsMaleMetalsMiddle AgedProsthesis FailureIonsMetals

Identifiers

PMID24847790
PMCPMC4105765
OpenAlexW1992625577

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.