Evidence map›Paper›PMID 24827970›Full record

ArticleInternational orthopaedics2014

The effect of a manufacturer recall on the threshold to revise a metal-on-metal hip.

Saket Tibrewal, Shiraz Sabah, Johann Henckel, Alister Hart

Open access · hybridAbstract read
PubMed Publisher
In one paragraph

Article in International orthopaedics, 2014. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed
1.2field-weighted citation impact, top 22% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed, 7 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Observational
  5. Article
  6. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors at 2 institutions in 1 country.

Saket TibrewalThe Royal National Orthopaedic Hospital, Stanmore, UK, sak11@hotmail.com.
Shiraz Sabah
Johann Henckel
Alister Hart
Royal National Orthopaedic Hospital · GBUniversity College London · GB

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeManagement of the unexplained, painful large diameter metal-on-metal (MOM) hip replacement is difficult. Although there are guidelines for surgeons, there is no clear documented evidence describing the overall threshold for revision surgery. The 2010 product recall of the DePuy Articular Surface Replacement (ASR) and subsequent media coverage may have increased patient and surgeon apprehension, resulting in earlier intervention, i.e. at a greater Oxford hip score (OHS) than expected. Our aim was to investigate whether the threshold for revision using known parameters was affected by the ASR recall. These parameters include poor clinical results (persistent pain or mechanical symptoms), pseudotumour or other progressive soft tissue involvement, osteolysis and high or rising metal ion levels.

methodsWe used our national referral database of MOM hips, which were revised between 2008 and 2012. Once inclusion and exclusion criteria were applied, we identified 240 patients--71 patients in the pre-recall group and 169 patients in the post-recall group.

resultsThe ASR product recall did not seem to affect the threshold for revision of a MOM hip, with no significant difference between the two groups in terms of the functional (median OHS = 17 pre-recall and 20 post-recall; p = 0.2109) and radiological (median inclination angle = 50 pre-recall and 48 post-recall; p = 0.3221) markers used to guide management. We did however discover that blood metal ion levels were higher in the post-recall group.

conclusionIssue of a product recall did not change the hip function threshold for revision surgery. The decision to revise a metal-on-metal hip is complex and should follow published guidelines, encompassing metal ion measurement and cross-sectional imaging where appropriate.

Indexed as

Hip ProsthesisMedical Device RecallsMetal-on-Metal Joint ProsthesesAdultAgedAged, 80 and overArthroplasty, Replacement, HipDatabases, FactualDecision MakingFemaleHumansMaleMiddle AgedReoperationYoung Adult

Identifiers

PMID24827970
OpenAlexW1988772527

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked