A novel removable shield attached to C-arm units against scattered X-rays from a patient's side.
Article in European radiology, 2014. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
5 citing papers in PubMed.
- Fundamental evaluation of a novel upright shield to protect medical personnel from scattered radiation during endoscopic retrograde cholangiopancreatography using a C-arm system.Journal of radiation research · 2026Article
- Strategy to Reduce the Collective Equivalent Dose for the Lens of the Physician's Eye Using Short Radiation Protection Curtains to Prevent Cataracts.Diagnostics (Basel, Switzerland) · 2021Article
- A multicenter study of radiation doses to the eye lenses of clinical physicians performing radiology procedures in Japan.Journal of occupational health · 2021Article
- Article
- Real-time, ray casting-based scatter dose estimation for c-arm x-ray system.Journal of applied clinical medical physics · 2017Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectivesWe invented a drape-like shield against scattered X-rays that can safely come into contact with medical equipment or people during fluoroscopically guided procedures. The shield can be easily removed from a C-arm unit using one hand. We evaluated the use of the novel removable shield during the endoscopic retrograde cholangiopancreatography (ERCP) procedure.
methodsWe measured the dose rate of scattered X-rays around endoscopists with and without this removable shield and surveyed the occupational doses to the ERCP staff. We also examined the endurance of the shield.
resultsThe removable shield reduced the dose rate of scattered X-rays to one-tenth and reduced the monthly dose to an endoscopist by at least two-fifths. For 2.5 years, there was no damage to the shield and no loosening of the seam. The bonding of the hook-and-loop fasteners did not weaken, although the powerful double-sided tapes made especially for plastic did.
conclusionsThe removable shield can reduce radiation exposure to the ERCP staff and may contribute to reducing the exposure to the eye lenses of operators. It would also be possible to expand its use to other fluoroscopically guided procedures besides ERCP because it is a light, simple, and useful device. KEY POINTS: • We invented a shield that can be removed from C-arm units with one hand. • The removable shield reduces the dose rate of X-rays to one-tenth. • The removable shield reduces operator exposure by two-fifths. • The removable shield is durable, lasting for several years. • The drape-like removable shield is light, simple, and useful.