Evidence map›Paper›PMID 24758340›Full record

ArticleJournal of cardiovascular electrophysiology2014

Compatibility of electroanatomical mapping systems with a concurrent percutaneous axial flow ventricular assist device.

Vaibhav R Vaidya, Christopher V Desimone, Malini Madhavan, Amit Noheria, Mohammed Shahid, Jacob Walters, Dorothy J Ladewig, Susan B Mikell, Susan B Johnson, Scott H Suddendorf and 1 more

Open access · greenAbstract read
In one paragraph

Article in Journal of cardiovascular electrophysiology, 2014. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 13 papers.

0numbers the graph read from it
0cells of the map it votes in
13citing papers in PubMed
0.6field-weighted citation impact, top 27% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

13 citing papers in PubMed, 20 citations in OpenAlex.

  1. Article
  2. Drug therapy and catheter ablation for management of arrhythmias in continuous flow left ventricular assist device's patients: a Clinical Consensus Statement of the European Heart Rhythm Association and the Heart Failure Association of the ESC.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2024
    Article
  3. Article
  4. Review
  5. Article
  6. Article
  7. [Application of hemodynamic optimization in the design of artificial heart].Sheng wu yi xue gong cheng xue za zhi = Journal of biomedical engineering = Shengwu yixue gongchengxue zazhi · 2020
    Article
  8. Comparison between His-bundle pacing guided by Ensite NavX system and conventional fluoroscopy.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2020
    Observational
  9. Article
  10. Article
  11. Article
  12. Article
  13. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

11 authors at 5 institutions in 1 country.

Vaibhav R VaidyaDepartment of Internal Medicine, Mayo Clinic, Rochester, Minnesota, USA.
Christopher V Desimone
Malini Madhavan
Amit Noheria
Mohammed Shahid
Jacob Walters
Dorothy J Ladewig
Susan B Mikell
Susan B Johnson
Scott H Suddendorf
Samuel J Asirvatham
Mayo Clinic · USMayo Clinic in Florida · USMayo Clinic in Arizona · USSt. Jude Medical Foundation · USUniversity of Pennsylvania · US

Funding

CARDIOVASOLOGYT32HL007111 · NHLBI · MAYO CLINIC ROCHESTER · PI Barry A. Borlaug · 1985 to 2026
$8.7M
NHLBI NIH HHS T32 HL007111
6 · The paper itself

Abstract

backgroundHemodynamic instability hinders activation and entrainment mapping during ventricular tachycardia ablation. The Impella 2.5 microaxial flow device (MFD; Abiomed Inc., Danvers, MA, USA) is used to prevent hemodynamic instability during electrophysiologic study. However, electromagnetic interference (EMI) generated by this device can preclude accurate electroanatomic mapping.

methodsImpella was placed in the left ventricle of 7 canines for circulatory support. Electroanatomic mapping during sinus rhythm, ventricular pacing, and ventricular fibrillation (VF) was performed using magnet- (CARTO3, Biosense Webster Inc., Diamond Bar, CA, USA) and impedance- (EnSite Velocity System/EnSite NavX, St. Jude Medical Inc., St. Paul, MN, USA) based systems. Distance from device to points with severe EMI precluding acquisition was compared to points with mild/no EMI. Two methods were used to reduce EMI: (1) titration of MFD performance, and (2) impedance-only mapping combined with manual annotation of activation.

resultsSevere EMI did not occur during impedance-based mapping. Severe EMI was observed using CARTO3 at 9.4% of all points attempted at maximum performance level (P8) of device. Severe EMI occurred at points closer to device (40.1 ± 16.8 mm) versus (55.5 ± 20.0 mm) for mild/no EMI, P < 0.0001. Severe EMI using CARTO3 was resolved by either (1) reduction of performance from P8 to P6 or (2) impedance-only mapping with manual annotation.

conclusionConcurrent use of MFD caused EMI to prevent acquisition of points with magnet-based mapping. Predictors for EMI were distance from device and performance level. Temporary reductions to P6 or impedance-only mapping are 2 methods to resolve EMI.

Indexed as

Heart-Assist DevicesVentricular Function, LeftAnimalsCardiac Pacing, ArtificialDisease Models, AnimalDogsElectric ImpedanceElectromagnetic FieldsElectrophysiologic Techniques, CardiacEquipment FailureHemodynamicsMalePredictive Value of TestsProsthesis DesignProsthesis ImplantationVentricular Fibrillationcatheter ablationelectroanatomical mapping systemsimpedanceImpellamagneticventricular tachycardia

Identifiers

PMID24758340
PMCPMC4447707
OpenAlexW2145634070

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked