The effect of metal artefact reduction on CT-based attenuation correction for PET imaging in the vicinity of metallic hip implants: a phantom study.

Roy Harnish, Sven Prevrhal, Abass Alavi, Habib Zaidi, Thomas F Lang

Open access · greenAbstract read
In one paragraph

Article in Annals of nuclear medicine, 2014. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
0.6field-weighted citation impact, top 33% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed, 12 citations in OpenAlex.

  1. Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 4 institutions in 5 countries.

Roy HarnishDepartment of Radiology and Biomedical Imaging, University of California, San Francisco, Box 0946, San Francisco, CA, 94143-0946, USA, Roy.Harnish@ucsf.edu.
Sven Prevrhal
Abass Alavi
Habib Zaidi
Thomas F Lang
University of California, San Francisco · USPhilips (Finland) · FIUniversity of Geneva · CHUniversity of Pennsylvania · US

Funding

FDG-PET Imaging in painful Hip ProsthesisR01AR048241 · NIAMS · UNIVERSITY OF PENNSYLVANIA · PI ALAVI, ABASS NONE · 2002 to 2005
$2.2M
Quantitative periprosthetic PET/CTR21AR055253 · NIAMS · UNIVERSITY OF CALIFORNIA, SAN FRANCISCO · PI LANG, THOMAS F · 2008 to 2009
$343k
NIAMS NIH HHS R01 AR048241NIAMS NIH HHS R01-AR048241NIAMS NIH HHS R21 AR055253NIAMS NIH HHS R21-AR055253
6 · The paper itself

Abstract

backgroundTo determine if metal artefact reduction (MAR) combined with a priori knowledge of prosthesis material composition can be applied to obtain CT-based attenuation maps with sufficient accuracy for quantitative assessment of (18)F-fluorodeoxyglucose uptake in lesions near metallic prostheses.

methodsA custom hip prosthesis phantom with a lesion-sized cavity filled with 0.2 ml (18)F-FDG solution having an activity of 3.367 MBq adjacent to a prosthesis bore was imaged twice with a chrome-cobalt steel hip prosthesis and a plastic replica, respectively. Scanning was performed on a clinical hybrid PET/CT system equipped with an additional external (137)Cs transmission source. PET emission images were reconstructed from both phantom configurations with CT-based attenuation correction (CTAC) and with CT-based attenuation correction using MAR (MARCTAC). To compare results with the attenuation-correction method extant prior to the advent of PET/CT, we also carried out attenuation correction with (137)Cs transmission-based attenuation correction (TXAC). CTAC and MARCTAC images were scaled to attenuation coefficients at 511 keV using a trilinear function that mapped the highest CT values to the prosthesis alloy attenuation coefficient. Accuracy and spatial distribution of the lesion activity was compared between the three reconstruction schemes.

resultsCompared to the reference activity of 3.37 MBq, the estimated activity quantified from the PET image corrected by TXAC was 3.41 MBq. The activity estimated from PET images corrected by MARCTAC was similar in accuracy at 3.32 MBq. CTAC corrected PET images resulted in nearly 40 % overestimation of lesion activity at 4.70 MBq. Comparison of PET images obtained with the plastic and metal prostheses in place showed that CTAC resulted in a marked distortion of the (18)F-FDG distribution within the lesion, whereas application of MARCTAC and TXAC resulted in lesion distributions similar to those observed with the plastic replica.

conclusionsMAR combined with a trilinear CT number mapping for PET attenuation correction resulted in estimates of lesion activity comparable in accuracy to that obtained with (137)Cs transmission-based attenuation correction, and far superior to estimates made without attenuation correction or with a standard CT attenuation map. The ability to use CT images for attenuation correction is a potentially important development because it obviates the need for a (137)Cs transmission source, which entails extra scan time, logistical complexity and expense.

Indexed as

ArtifactsHip ProsthesisMetalsCesium RadioisotopesCobaltFluorodeoxyglucose F18Phantoms, ImagingPlasticsPositron-Emission TomographyTomography, X-Ray ComputedCesium RadioisotopesCobaltFluorodeoxyglucose F18MetalsPlastics

Identifiers

PMID24710757
PMCPMC4101148
OpenAlexW1970242536

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.