ArticleEuropean spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society2014
The effect of design parameters of interspinous implants on kinematics and load bearing: an in vitro study.
Article in European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society, 2014. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
8 citing papers in PubMed, 15 citations in OpenAlex.
- Standalone interspinous process devices for lumbar spinal stenosis: a critical literature review.BMC surgery · 2026Review
- The effect of lifting load on the kinematic characteristics of lumbar spinous process in vivo.Surgical and radiologic anatomy : SRA · 2023Article
- Topping-off surgery vs posterior lumbar interbody fusion for degenerative lumbar disease: a comparative study of clinical efficacy and adjacent segment degeneration.Journal of orthopaedic surgery and research · 2019Article
- Influence of Deviated Centers of Rotation on Kinematics and Kinetics of a Lumbar Functional Spinal Unit: An In Vitro Study.Medical science monitor : international medical journal of experimental and clinical research · 2019Article
- Prediction of the Spinal Musculoskeletal Loadings during Level Walking and Stair Climbing after Two Types of Simulated Interventions in Patients with Lumbar Disc Herniation.Journal of healthcare engineering · 2019Article
- Radioisotopic assessment of bone metabolism of the operated vertebra after inter-process stabilizer implantation in the lumbar segment of the spine.Archives of medical science : AMS · 2017Article
- Change of sagittal spinal alignment and its association with pain and function after lumbar surgery augmented with an interspinous implant.Scoliosis and spinal disorders · 2017Article
- Interspinous dynamic stabilization adjacent to fusion versus double-segment fusion for treatment of lumbar degenerative disease with a minimum follow-up of three years.International orthopaedics · 2016Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
introductionA number of concepts with controversy approaches are currently discussed for interspinous stabilization (IPS). However, comparative biomechanical studies among the different systems are rare. Nevertheless, it remains unclear which biomechanical characteristics are influenced by different design features of these implants, such as implant stiffness or an additional tension band. Therefore, the aim of the present study was to compare different interspinous implants to investigate the biomechanical impact of IPS implant design on intersegmental kinematics, such as range of motion, neutral zone, center of rotation (COR), as well as load transfer like intradiscal pressure (IDP), to gain additional experience for clinical indications and limitations. MATERIAL AND
methodTwelve human lumbar spine specimens were tested in a spine loading apparatus. In vitro flexibility testing was performed by applying pure bending moments of 7.5 Nm without and with additional preload of 400 N in the three principal motion planes. Four interspinous implants, Coflex "COF" (Paradigm Spine, Germany), Wallis "WAL" (Abbott Laboratories, France), DIAM "DIA" (Sofamor Danek, France) and InterActiv (Aesculap AG, Germany) with two treatment options (without dorsal tensioning "IAO" and with dorsal tensioning "IAM") were consecutively tested in comparison to the native situation "NAT" and to a defect situation "DEF" of the functional spinal unit. The tested IPS devices are comprised of a compression stiffness range of 133 to 1,674 N/mm and a tensile stiffness range of 0-39 N/mm. Range of motion, neutral zone, center of rotation and intradiscal pressure were analyzed for all instrumentation steps and load cases.
conclusionFor the IPS, we found a correlation between compression stiffness and stabilization in extension. Here, the system with the lowest stiffness, DIA, displayed nearly no stabilization of the treated segment, whereas the system with the highest stiffness, WAL and COF, was most pronounced. This applies also for the correlation between device stiffness and IDP. In flexion only the degree of stabilization is in correlation with the tensile stiffness, whereas the IDP stays constant and is not affected by the different tensile stiffness. IPS is not able to stabilize in the frontal and transversal plane. Furthermore IPS does not substantially alter the location of the COR.
Indexed as