ArticleJournal of neurointerventional surgery2015
Quantitative analysis of high-resolution, contrast-enhanced, cone-beam CT for the detection of intracranial in-stent hyperplasia.
Article in Journal of neurointerventional surgery, 2015. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 14 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
14 citing papers in PubMed, 35 citations in OpenAlex.
- Evaluation of FeMnN alloy bioresorbable flow diverting stents in the rabbit abdominal aorta.Bioactive materials · 2025Article
- Intracranial Aneurysms Treated with a Novel Coated Low-Profile Flow Diverter (p48 HPC)-A Single-Center Experience and an Illustrative Case Series.Brain sciences · 2025Article
- Evaluation of FeMnN alloy bioresorbable flow diverters in the rabbit elastase induced aneurysm model.Frontiers in bioengineering and biotechnology · 2025Article
- Benchtop proof of concept and comparison of iron- and magnesium-based bioresorbable flow diverters.Journal of neurosurgery · 2023Article
- Medical Imaging Compatibility of Magnesium- and Iron-Based Bioresorbable Flow Diverters.AJNR. American journal of neuroradiology · 2023Article
- Intimal hyperplasia detection after aneurysm treatment by flow diversion using magnetic resonance vessel wall imaging.The neuroradiology journal · 2022Article
- Identification of Carotid Artery Microstructure and Plaque Rupture Using C-Arm Cone-Beam CT: A Case Report.Frontiers in neurology · 2021Article
- Filament inadequate wall apposition of the different ends of flow diverters in the abdominal aorta of rabbits.The neuroradiology journal · 2020Article
- High frequency optical coherence tomography assessment of homogenous neck coverage by intrasaccular devices predicts successful aneurysm occlusion.Journal of neurointerventional surgery · 2019Article
- Optimizing contrast-enhanced cone-beam CT protocol to facilitate simultaneous visualization of neurovascular pathologies and surrounding structures of interest.Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences · 2019Article
- In-stent Stenosis after p64 Flow Diverter Treatment.Clinical neuroradiology · 2018Article
- Open-cell stent and use of cone-beam CT enables a safe and effective coil embolization of true ophthalmic artery and anterior choroidal artery aneurysms with preservation of parent vessel: Clinical and angiographic results.Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences · 2018Article
- Use of self-expanding stents for better intracranial flow diverter wall apposition.Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences · 2017Article
- Intravenous C-Arm Conebeam CT Angiography following Long-Term Flow-Diverter Implantation: Technologic Evaluation and Preliminary Results.AJNR. American journal of neuroradiology · 2016Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundIntracranial in-stent hyperplasia is a stroke-associated complication that requires routine surveillance.
objectiveTo compare the results of in vivo experiments to determine the accuracy and precision of in-stent hyperplasia measurements obtained with modified C-arm contrast-enhanced, cone-beam CT (CE-CBCT) imaging with those obtained by 'gold standard' histomorphometry. Additionally, to carry out clinical analyses comparing this CE-CBCT protocol with digital subtraction angiography (DSA).
methodsA non-binned CE-CBCT protocol (VasoCT) was used that acquires x-ray images with a small field-of-view and applies a full-scale reconstruction algorithm providing high-resolution three-dimensional (3D) imaging with 100 µm isotropic voxels. In an vivo porcine model, VasoCT cross-sectional area measurements were compared with gold standard vessel histology. VasoCT and DSA were used to calculate in-stent stenosis in 23 imaging studies.
resultsPorcine VasoCT cross-sectional stent, lumen, and in-stent hyperplasia areas strongly correlated with histological measurements (r(2)=0.97, 0.93, 0.90; slope=1.14, 1.07, and 0.76, respectively; p<0.0001). Clinical VasoCT percentage stenosis correlated well with DSA percentage stenosis (r(2)=0.84; slope=0.76), and the two techniques were free of consistent bias (Bland-Altman, bias=3.29%; 95% CI -14.75% to 21.33%). An illustrative clinical case demonstrated the advantages of VasoCT, including 3D capability and non-invasive IV contrast administration, for detection of in-stent hyperplasia.
conclusionsC-arm VasoCT is a high-resolution 3D capable imaging technique that has been validated in an animal model for measurement of in-stent tissue growth. Successful clinical implementation of the protocol was performed in a small case series.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials