Impact of partial pressure of oxygen in blood samples on the performance of systems for self-monitoring of blood glucose.
Article in Diabetes technology & therapeutics, 2014. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 10 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
10 citing papers in PubMed, 19 citations in OpenAlex.
- A universal oxygen scavenger for oxidase-based biosensors.Science advances · 2025Article
- A direct effect of the hematocrit on blood glucose: Evidence from hypoxia- and erythropoietin-treated mice.Science advances · 2025Article
- Clinical Accuracy of a Glucose Oxidase-Based Blood Glucose Test-Strip Across Extremes of Oxygen Partial Pressure.Journal of diabetes science and technology · 2024Article
- Distribution of Oxygen Partial Pressures in Native Capillary Fingertip Blood Samples From a Large Diabetes Patient Population.Journal of diabetes science and technology · 2023Article
- Performance evaluation of three i-SENS glucometers using arterial blood samples compared with the YSI 2300 Glucose Analyzer.Journal of clinical laboratory analysis · 2020Article
- Proof of Concept Study to Assess the Influence of Oxygen Partial Pressure in Capillary Blood on SMBG Measurements.Journal of diabetes science and technology · 2019Article
- Pilots and Diabetes Technology.Journal of diabetes science and technology · 2017Article
- Self-monitoring of blood glucose: impact of a time delay between capillary blood sampling and glucose measurement.Journal of diabetes science and technology · 2014Article
- Influence of partial pressure of oxygen in blood samples on measurement performance in glucose-oxidase-based systems for self-monitoring of blood glucose.Journal of diabetes science and technology · 2013Article
- Partial pressure of oxygen in capillary blood samples from the fingertip.Journal of diabetes science and technology · 2013Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundThe partial pressure of oxygen (pO2) in blood samples can affect glucose measurements with oxygen-sensitive systems. In this study, we assessed the influence of different pO2 levels on blood glucose (BG) measurements with five glucose oxidase (GOD) systems and one glucose dehydrogenase (GDH) system. All selected GOD systems were indicated by the manufacturers to be sensitive to increased oxygen content of the blood sample. MATERIALS AND
methodsVenous blood samples of 16 subjects (eight women, eight men; mean age, 52 years; three with type 1 diabetes, four with type 2 diabetes, and nine without diabetes) were collected. Aliquots of each sample were adjusted to the following pO2 values: ≤45 mm Hg, approximately 70 mm Hg, and ≥150 mm Hg. For each system, five consecutive measurements on each sample were performed using the same test strip lot. Relative differences between the mean BG value at a pO2 level of approximately 70 mm Hg, which was considered to be similar to pO2 values in capillary blood samples, and the mean BG value at pO2 levels ≤45 mm Hg and ≥150 mm Hg were calculated.
resultsThe GOD systems showed mean relative differences between 11.8% and 44.5% at pO2 values ≤45 mm Hg and between -14.6% and -21.2% at pO2 values ≥150 mm Hg. For the GDH system, the mean relative differences were -0.3% and -0.2% at pO2 values ≤45 mm Hg and ≥150 mm Hg, respectively.
conclusionsThe magnitude of the pO2 impact on BG measurements seems to vary among the tested oxygen-sensitive GOD systems. The pO2 range in which oxygen-sensitive systems operate well should be provided in the product information.