Analytic evaluation of a new glucose meter system in 15 different critical care settings.
Article in Journal of diabetes science and technology, 2013. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 11 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
11 citing papers in PubMed.
- Accuracy of blood glucose measurements using the NOVA StatStrip® glucometer during cardiac surgery: a prospective observational study.Canadian journal of anaesthesia = Journal canadien d'anesthesie · 2019Trial
- Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus.Diabetes care · 2023Review
- Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus.Clinical chemistry · 2023Article
- Unintended Consequence of High-Dose Vitamin C Therapy for an Oncology Patient: Evaluation of Ascorbic Acid Interference With Three Hospital-Use Glucose Meters.Journal of diabetes science and technology · 2021Article
- A comparison of the glycemic effects of glucagon using two dose ranges in neonates and infants with hypoglycemia.Journal of perinatology : official journal of the California Perinatal Association · 2020Article
- Comparison of Venous and Capillary Sampling in Oral Glucose Testing for the Diagnosis of Gestational Diabetes Mellitus: A Diagnostic Accuracy Cross-Sectional Study Using Accu-Chek Inform II.Diagnostics (Basel, Switzerland) · 2020Article
- Accuracy of Capillary and Arterial Whole Blood Glucose Measurements Using a Glucose Meter in Patients under General Anesthesia in the Operating Room.Anesthesiology · 2017Article
- Pre-morbid glycemic control modifies the interaction between acute hypoglycemia and mortality.Intensive care medicine · 2016Observational
- Impact of Glucose Meter Error on Glycemic Variability and Time in Target Range During Glycemic Control After Cardiovascular Surgery.Journal of diabetes science and technology · 2015Article
- Article
- Analysis: New point-of-care blood glucose monitoring system for the hospital demonstrates satisfactory analytical accuracy using blood from critically ill patients--an important step toward improved blood glucose control in the hospital.Journal of diabetes science and technology · 2013Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundMaintaining appropriate glycemic control in critically ill patients reduces morbidity and mortality. The use of point-of-care (POC) glucose devices is necessary to obtain rapid results at the patient's bedside. However, the devices should be thoroughly tested in the intended population before implementation. The use of POC glucose meters in critically ill patients has been questioned both in the literature and by regulatory agencies. The aim of this study was to determine if the ACCU-CHEK® Inform II system (Roche Diagnostics) POC glucose meter demonstrated the desired accuracy and precision, as defined by Clinical and Laboratory Standards Institute guideline POCT12-A3, in a large number of critically ill patients from multiple intensive care settings at two academic medical centers.
methodsA total of 1200 whole blood meter results from 600 patients were compared with central laboratory plasma values. Whole blood aliquots from venous samples were used to obtain duplicate meter results with the remaining sample being processed to obtain plasma for central laboratory testing within 5 min of meter testing.
resultsA total of 1185 (98.8%) of the new meter's glucose values were within ± 12.5% (± 12 mg/dl for values ≥ 100 mg/dl) of the comparative laboratory glucose values, and 1198 (99.8%) were within ± 20% (± 20 mg/dl for values <100 mg/dl).
conclusionsConsidering the large number of patients from numerous critical care units examined, the new glucose meter system appears to have sufficient analytic accuracy for use in critically ill patients.