Evidence map›Paper›PMID 23917990›Full record

ArticleClinical orthopaedics and related research2014

The effect of geometric variations in posterior-stabilized knee designs on motion characteristics measured in a knee loading machine.

Peter S Walker, Michael T Lowry, Anoop Kumar

Abstract read
In one paragraph

Article in Clinical orthopaedics and related research, 2014. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Motion type and knee articular conformity influenced mid-flexion stability of a single radius knee prosthesis.Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA · 2019
    Article
  5. Article
  6. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors.

Peter S WalkerDepartment of Orthopaedic Surgery, New York University-Hospital for Joint Diseases, New York, NY, USA, Peter.Walker@nyumc.org.
Michael T Lowry
Anoop Kumar

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundIn different posterior-stabilized (PS) total knees, there are considerable variations in condylar surface radii and cam-post geometry. To what extent these variations affect kinematics is not known. Furthermore, there are no clearly defined ideal kinematics for a total knee. QUESTIONS/PURPOSES: The purposes of this study were to determine (1) what the kinematic differences are caused by geometrical variations between PS total knee designs in use today; and (2) what design characteristics will produce kinematics that closely resemble that of the normal anatomic knee.

methodsFour current PS designs with different geometries and one experimental asymmetric PS design, with a relatively conforming medial side, were tested in a purpose-built machine. The machine applied combinations of compressive, shear, and torque forces at a sequence of flexion angles to represent a range of everyday activities, consistent with the ASTM standard test for measuring constraint. The femorotibial contact points, the neutral path of motion, and the AP and internal-external laxities were used as the kinematic indicators.

resultsThe PS designs showed major differences in motion characteristics among themselves and with motion data from anatomic knees determined in a previous study. Abnormalities in the current designs included symmetric mediolateral motion, susceptibility to excessive AP medial laxity, and reduced laxity in high flexion. The asymmetric-guided motion design alleviated some but not all of the abnormalities.

conclusionsCurrent PS designs showed kinematic abnormalities to a greater or lesser extent. An asymmetric design may provide a path to achieving a closer match to anatomic kinematics. CLINICAL RELEVANCE: One criterion for the evaluation of PS total knees is how closely the kinematics of the prosthesis resemble that of the anatomic knee, because this is likely to affect the quality of function.

Indexed as

Arthroplasty, Replacement, KneeKnee ProsthesisProsthesis DesignBiomechanical PhenomenaHumansKneeKnee JointRange of Motion, Articular

Identifiers

PMID23917990
PMCPMC3889438

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.