Evidence map›Paper›PMID 23913308›Full record

ArticleThe British journal of radiology2013

Cone beam CT guidance provides superior accuracy for complex needle paths compared with CT guidance.

W M H Busser, S J Braak, J J Fütterer, M J L van Strijen, Y L Hoogeveen, F de Lange, L J Schultze Kool

Abstract readComparative Study
In one paragraph

Article in The British journal of radiology, 2013. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 24 papers.

0numbers the graph read from it
0cells of the map it votes in
24citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

24 citing papers in PubMed.

  1. Review
  2. Review
  3. Review
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Article
  10. Review
  11. Review
  12. Article
  13. Article
  14. Article
  15. Article
  16. Article
  17. Review
  18. Review
  19. Article
  20. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

W M H BusserDepartment of Radiology, Radboud University Nijmegen Medical Centre, Nijmegen, Netherlands. w.busser@rad.umcn.nl
S J Braak
J J Fütterer
M J L van Strijen
Y L Hoogeveen
F de Lange
L J Schultze Kool

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectiveTo determine the accuracy of cone beam CT (CBCT) guidance and CT guidance in reaching small targets in relation to needle path complexity in a phantom.

methodsCBCT guidance combines three-dimensional CBCT imaging with fluoroscopy overlay and needle planning software to provide real-time needle guidance. The accuracy of needle positioning, quantified as deviation from a target, was assessed for inplane, angulated and double angulated needle paths. Four interventional radiologists reached four targets along the three paths using CBCT and CT guidance. Accuracies were compared between CBCT and CT for each needle path and between the three approaches within both modalities. The effect of user experience in CBCT guidance was also assessed.

resultsAccuracies for CBCT were significantly better than CT for the double angulated needle path (2.2 vs 6.7 mm, p<0.001) for all radiologists. CBCT guidance showed no significant differences between the three approaches. For CT, deviations increased with increasing needle path complexity from 3.3 mm for the inplane placements to 4.4 mm (p=0.007) and 6.7 mm (p<0.001) for the angulated and double angulated CT-guided needle placements, respectively. For double angulated needle paths, experienced CBCT users showed consistently higher accuracies than trained users [1.8 mm (range 1.2-2.2) vs 3.3 mm (range 2.1-7.2) deviation from target, respectively; p=0.003].

conclusionIn terms of accuracy, CBCT is the preferred modality, irrespective of the level of user experience, for more difficult guidance procedures requiring double angulated needle paths as in oncological interventions. ADVANCES IN KNOWLEDGE: Accuracy of CBCT guidance has not been discussed before. CBCT guidance allows accurate needle placement irrespective of needle path complexity. For angulated and double-angulated needle paths, CBCT is more accurate than CT guidance.

Indexed as

NeedlesClinical CompetenceCone-Beam Computed TomographyFluoroscopyHumansImaging, Three-DimensionalPhantoms, ImagingSoftwareTomography, X-Ray Computed

Identifiers

PMID23913308
PMCPMC3798334

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.