Evidence map›Paper›PMID 23643546›Full record

ArticleThe Annals of thoracic surgery2013

Biventricular finite element modeling of the Acorn CorCap Cardiac Support Device on a failing heart.

Jonathan F Wenk, Liang Ge, Zhihong Zhang, Dimitri Mojsejenko, D Dean Potter, Elaine E Tseng, Julius M Guccione, Mark B Ratcliffe

Open access · greenAbstract readComparative Study
In one paragraph

Article in The Annals of thoracic surgery, 2013. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.

0numbers the graph read from it
0cells of the map it votes in
8citing papers in PubMed
3.5field-weighted citation impact, top 8% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

8 citing papers in PubMed, 30 citations in OpenAlex.

  1. Review
  2. Article
  3. Review
  4. Article
  5. Review
  6. Article
  7. Article
  8. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 4 institutions in 1 country.

Jonathan F WenkDepartment of Mechanical Engineering, University of Kentucky, Lexington, Kentucky, USA.
Liang Ge
Zhihong Zhang
Dimitri Mojsejenko
D Dean Potter
Elaine E Tseng
Julius M Guccione
Mark B Ratcliffe
San Francisco VA Medical Center · USUniversity of California, San Francisco · USUniversity of Kentucky · USMayo Clinic · US

Funding

QUANTITATIVE ANALYSIS OF LEFT VENTRICULPLASTY--CHFR01HL063348 · NHLBI · NORTHERN CALIFORNIA INSTITUTE RES &EDUC · PI RATCLIFFE, MARK B · 2000 to 2017
$6.8M
VIRTUAL TOOLS FOR CARDIAC VENTRICULAR REMODELING SURGERYR01HL077921 · NHLBI · UNIVERSITY OF CALIFORNIA, SAN FRANCISCO · PI GUCCIONE, JULIUS MATTEO · 2006 to 2015
$3.4M
NHLBI NIH HHS 5R01 HL063348NHLBI NIH HHS 5R01 HL077921NHLBI NIH HHS R01 HL063348NHLBI NIH HHS R01 HL077921
6 · The paper itself

Abstract

backgroundThe Acorn CorCap Cardiac Support Device (CSD; Acorn Cardiovascular Inc, St. Paul, MN) is a woven polyester jacket that is placed around the heart and designed to reverse the progressive remodeling associated with dilated cardiomyopathy. However, the effects of the Acorn CSD on myofiber stress and ventricular function remain unknown. We tested the hypothesis that the Acorn CSD reduces end-diastolic (ED) myofiber stress.

methodsA previously described weakly coupled biventricular finite element (FE) model and circulatory model based on magnetic resonance images of a dog with dilated cardiomyopathy was used. Virtual applications of the CSD alone (Acorn), CSD with rotated fabric fiber orientation (rotated), CSD with 5% prestretch (tight), and CSD wrapped only around the left ventricle (LV; LV-only) were performed, and the effect on myofiber stress at ED and pump function was calculated.

resultsThe Acorn CSD has a large effect on ED myofiber stress in the LV free wall, with reductions of 55%, 79%, 92%, and 40% in the Acorn, rotated, tight, and LV-only cases, respectively. However, there is a tradeoff in which the Acorn CSD reduces stroke volume at LV end-diastolic pressure of 8 mm Hg by 23%, 25%, 30%, and 7%, respectively, in the Acorn, rotated, tight, and LV-only cases.

conclusionsThe Acorn CSD significantly reduces ED myofiber stress. However, CSD wrapped only around the LV was the only case with minimal negative effect on pump function. Findings suggest that LV-only CSD and Acorn fabric orientation should be optimized to allow maximal myofiber stress reduction with minimal reduction in pump function.

Indexed as

Finite Element AnalysisHeart-Assist DevicesAnimalsCardiomyopathy, DilatedDisease Models, AnimalDogsEquipment DesignEquipment SafetyHeart Function TestsMaterials TestingRandom AllocationSensitivity and SpecificityStress, MechanicalVentricular Remodeling

Identifiers

PMID23643546
PMCPMC3702266
OpenAlexW2118013368

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.