Evidence map›Paper›PMID 23538407›Full record

ArticleAJNR. American journal of neuroradiology

Histopathologic evaluation of arterial wall response to 5 neurovascular mechanical thrombectomy devices in a swine model.

B Gory, D Bresson, I Kessler, M L Perrin, A Guillaudeau, K Durand, S Ponsonnard, C Couquet, C Yardin, C Mounayer

Open access · hybridAbstract readComparative StudyEvaluation Study
In one paragraph

Article in AJNR. American journal of neuroradiology. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 42 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
42citing papers in PubMed, 1 pooled it
4.7field-weighted citation impact, top 5% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

42 citing papers in PubMed, 1 synthesis or guideline pooled it, 108 citations in OpenAlex.

  1. Large animals in neurointerventional research: A systematic review on models, techniques and their application in endovascular procedures for stroke, aneurysms and vascular malformations.Journal of cerebral blood flow and metabolism : official journal of the International Society of Cerebral Blood Flow and Metabolism · 2019
    Pooled it
  2. Adhesion of vessel wall to stentriever during combined technique for mechanical thrombectomy in acute ischemic stroke: A histomorphological study.Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences · 2026
    Article
  3. Is first-pass effect a meaningful metric to evaluate thrombectomy technologies?Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences · 2025
    Article
  4. Article
  5. Article
  6. Article
  7. Review
  8. Article
  9. Forced suction thrombectomy in patients with acute ischemic stroke using the SOFIA Plus device.Journal of cerebrovascular and endovascular neurosurgery · 2022
    Article
  10. Article
  11. Article
  12. Observational
  13. Reperfusion therapy in a patient with early recurrence of ischemic stroke.Neurological sciences : official journal of the Italian Neurological Society and of the Italian Society of Clinical Neurophysiology · 2022
    Article
  14. Article
  15. Article
  16. Review
  17. Article
  18. Article
  19. Interventional stroke treatment - Is it also safe for arteries? Looking at thrombectomy wall damage through clot histology.Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences · 2021
    Article
  20. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

10 authors at 2 institutions in 2 countries.

B GoryDepartments of Interventional Neuroradiology.
D Bresson
I Kessler
M L Perrin
A Guillaudeau
K Durand
S Ponsonnard
C Couquet
C Yardin
C Mounayer
Hôpital Universitaire Dupuytren · FRUniversidade de Brasília · BR

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

background and purposeFive commercial devices are available for mechanical thrombectomy in acute ischemic stroke. This study evaluated and compared the resultant arterial damage from these devices. MATERIALS AND

methodsWall damage after 4 wall-contact devices (the Merci retriever, Catch thromboembolectomy system, and Solitaire FR revascularization devices of 4 and 6 mm) and 1 aspiration device (the Penumbra System) was evaluated in the superficial femoral arteries of 20 male swine. Each device was tested with and without intraluminal clot. Twenty control vessels were not subjected to any intervention. Acute histopathologic changes were evaluated.

resultsIn the device samples, endothelial denudation (72.8 ± 29.4% versus 0.9 ± 1.9%, P < .0001), medial layer edema (52 ± 35.9% versus 18.1 ± 27.8%, P = .004), and mural thrombus (5.3 ± 14.2% versus 0%, P = .05) were found to a greater extent compared with the control samples. The aspiration device provoked more intimal layer (100 ± 79.1% versus 58.8 ± 48.9%, P = .27) and medial layer (75 ± 35.4% versus 46.3 ± 34.8%, P = .13) edema than the wall-contact devices.

conclusionsAll devices caused vascular injuries extending into the medial layer. The aspiration device was associated with more intimal and medial layer edema, compared with the wall-contact devices except for the Catch thromboembolectomy system.

Indexed as

AnimalsEdemaEquipment DesignEquipment Failure AnalysisFemoral ArteryMaleMechanical ThrombolysisPeripheral Arterial DiseaseSwine

Identifiers

PMID23538407
PMCPMC7964832
OpenAlexW2166331456

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.