Evidence map›Paper›PMID 22915946›Full record

ArticleMedical devices (Auckland, N.Z.)2011

Validation of the BPLab(®) 24-hour blood pressure monitoring system according to the European standard BS EN 1060-4:2004 and British Hypertension Society protocol.

Sergey A Koudryavtcev, Vyacheslav M Lazarev

RetractedOpen access · goldAbstract readRetracted Publication
In one paragraph

Article in Medical devices (Auckland, N.Z.), 2011. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It has been retracted, and should not be counted. Cited by 14 papers.

0numbers the graph read from it
0cells of the map it votes in
14citing papers in PubMed
1.0field-weighted citation impact, top 21% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

14 citing papers in PubMed, 21 citations in OpenAlex.

  1. Review
  2. Article
  3. Observational
  4. Article
  5. Review
  6. Article
  7. Tendency for age-specific mortality with hypertension in the European Union from 1980 to 2011.International journal of clinical and experimental medicine · 2015
    Article
  8. Article
  9. Article
  10. Article
  11. Day-to-day repeatability of the Pulse Time Index of Norm.Medical devices (Auckland, N.Z.) · 2014
    Article
  12. Article
  13. Article
  14. Article
4 · The record

Corrections and comments

5 · Who and what money

Authors and funding

2 authors at 1 institution in 1 country.

Sergey A KoudryavtcevDepartment of Military Medicine, Municipal Medical and Preventive Treatment Facility, 'Clinical Hospital No 12', Sormovsky District, Nizhny Novgorod, Russian Federation.
Vyacheslav M Lazarev
Privolzhsky Research Medical University · RU

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

introductionAutomatic blood pressure (BP) measuring devices are more and more often used in BP self-checks and in 24-hour BP monitoring. Nowadays, 24-hour BP monitoring is a necessary procedure in arterial hypertension treatment. The aim of this study was to validate the BPLab(®) ambulatory blood pressure monitor according to the European standard BS EN 1060-4:2004 and the British Hypertension Society (BHS) protocol, as well as to work out solutions regarding the suitability of using this device in clinical practice.

methodsA group of 85 patients of both sexes and different ages, who voluntarily agreed to take part in the tests and were given detailed instructions on the measurement technique were recruited for this study. The results of the BP measurement obtained by a qualified operator using the BPLab(®) device were compared with the BP values measured using the Korotkov auscultatory method. Data were obtained simultaneously by two experts with experience of over 10 years and had completed a noninvasive BP measurement standardization training course. Discrepancies in the systolic and diastolic BP measurements (N = 510; 255 for each expert) were analyzed according to the criteria specified in the BHS-93 protocol.

resultsThe device passed the requirements of the European Standard BS EN 1060-4:2004 and was graded 'A' according to the criteria of the BHS protocol for both systolic BP and diastolic BP.

conclusionThe BPLab(®) 24-hour ambulatory blood pressure monitoring device may be recommended for extensive clinical use.