ArticleAJNR. American journal of neuroradiology2013
Experimental evaluation of immediate recanalization effect and recanalization efficacy of a new thrombus retriever for acute stroke treatment in vivo.
Article in AJNR. American journal of neuroradiology, 2013. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 13 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
13 citing papers in PubMed, 27 citations in OpenAlex.
- Review
- First-Pass Reperfusion by Mechanical Thrombectomy in Acute M1 Occlusion: The Size of Retriever Matters.Frontiers in neurology · 2021Article
- Relevance of Porcine Stroke Models to Bridge the Gap from Pre-Clinical Findings to Clinical Implementation.International journal of molecular sciences · 2020Review
- Fast Stent Retrieval Improves Recanalization Rates of Thrombectomy: Experimental Study on Different Thrombi.AJNR. American journal of neuroradiology · 2020Article
- Imaging Clot Characteristics in Stroke and its Possible Implication on Treatment.Clinical neuroradiology · 2020Review
- Preclinical Evaluation of the NeVaTM Stent Retriever: Safety and Efficacy in the Swine Thrombectomy Model.Interventional neurology · 2018Article
- Retrograde semi-retrieval technique for combined stentriever plus aspiration thrombectomy in acute ischemic stroke.Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences · 2017Article
- Experimental evaluation of stent retrievers' mechanical properties and effectiveness.Journal of neurointerventional surgery · 2017Article
- The pREset Stent Retriever for Endovascular Treatment of Stroke Caused by MCA Occlusion: Safety and Clinical Outcome.Clinical neuroradiology · 2016Article
- Flat panel detector computed tomography for the interaction between contrast-enhanced thrombi and stent retrievers in stroke therapy: a pilot study.Clinical neuroradiology · 2014Article
- Clinical experience with the pREset stent retriever for the treatment of acute ischemic stroke--a review of 271 consecutive cases.Neuroradiology · 2014Article
- Endovascular treatment of acute ischemic stroke.Journal of neurosciences in rural practice · 2013Review
- Mechanical thrombectomy devices for endovascular management of acute ischemic stroke: Duke stroke center experience.Asian journal of neurosurgery · 2012Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
background and purposeCurrently, several new stent retriever devices for acute stroke treatment are under development and early clinical evaluation. Preclinical testing under standardized conditions is an important first step to evaluate the technical performance and potential of these devices. The aim of this study was to evaluate the immediate recanalization effect, recanalization efficacy, thrombus-device interaction, and safety of a new stent retriever intended for thrombectomy in patients with acute stroke. MATERIAL AND
methodsThe pREset thrombectomy device (4 × 20 mm) was evaluated in 16 vessel occlusions in an established swine model. Radiopaque thrombi (10-mm length) were used for visualization of thrombus-device interaction during application and retrieval. Flow-restoration effect immediately after deployment and after 5-minute embedding time before retrieval, recanalization rate after retrieval, thromboembolic events, and complications were assessed. High-resolution FPCT was performed to illustrate thrombus-device interaction during the embedding time.
resultsImmediate flow restoration was achieved in 75% of occlusions. An increase or stable percentage of recanalizations during embedding time before retrieval was seen in 56.3%; a decrease, in 12.5%; reocclusion of a previously recanalized vessel, in 18.8%; and no recanalization effect at all, in 12.5%. Complete recanalization (TICI 3) after retrieval was achieved in 93.8%; partial recanalization (TICI 2b), in 6.2%. No distal thromboembolic events were observed. High-resolution FPCT illustrated entrapment of the thrombus between the stent struts and compression against the contralateral vessel wall, leading to partial flow restoration. During retrieval, the thrombus was retained in a straight position within the stent struts.
conclusionsIn this experimental study, the pREset thrombus retriever showed a high recanalization rate in vivo. High-resolution FPCT allows detailed illustration of the thrombus-device interaction during embedding time and is advocated as an add-on tool to the animal model used in this study.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials