Evidence map›Paper›PMID 22833262›Full record

ArticleSurgical endoscopy2013

Benchtop evaluation of pressure barrier insufflator and standard insufflator systems.

Kenneth G Nepple, Dorina Kallogjeri, Sam B Bhayani

Abstract readEvaluation Study
PubMed Publisher
In one paragraph

Article in Surgical endoscopy, 2013. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 23 papers, 3 of them syntheses that pooled it.

0numbers the graph read from it
0cells of the map it votes in
23citing papers in PubMed, 3 pooled it
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

23 citing papers in PubMed, 3 syntheses or guidelines pooled it.

  1. Pooled it
  2. Pooled it
  3. Pooled it
  4. Trial
  5. Trial
  6. Trial
  7. Article
  8. Radial fan-based COSurgical endoscopy · 2026
    Article
  9. Article
  10. Article
  11. Article
  12. Article
  13. Article
  14. The impact of AirSealWorld journal of urology · 2023
    Article
  15. Article
  16. Article
  17. Article
  18. Review
  19. Article
  20. Advances in laparoscopic surgery in urology.Nature reviews. Urology · 2016
    Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors.

Kenneth G NeppleDivision of Urology, Department of Surgery, Washington University School of Medicine, St. Louis, MO 63110, USA. nepplek@wustl.edu
Dorina Kallogjeri
Sam B Bhayani

Funding

WASHINGTON UNIVERSITY CER MENTORED CAREER DEVELOPMENT AWARD KM1KM1CA156708 · NCI · WASHINGTON UNIVERSITY · PI FRASER, VICTORIA J. · 2010 to 2010
$2.5M
NCI NIH HHS KM1 CA156708
6 · The paper itself

Abstract

backgroundPrevious experimental research has reported minimal differences in pressure maintenance between different versions of standard insufflators (SI). However, a recent report identified potential clinical benefits with a valveless pressure barrier insufflator (PBI). We sought to perform a benchtop objective evaluation of SI and PBI systems.

methodsA rigid box system with continuous pressure manometry was used to evaluate a PBI (Surgiquest Airseal) and two SIs (SI1 = Stryker PneumoSure High Flow Insufflator and SI2 = Storz SCB Thermoflator). Pressure maintenance of 15 mmHg was evaluated during experimental conditions of leakage from a 5 mm port site, leakage from a 12 mm port site, and continuous suction.

resultsWith leakage from the 5 mm port site, the PBI maintained pressure of >13 mmHg whereas the pressures dropped moderately with the SI1 (7-13 mmHg) and SI2 insufflators (3-7 mmHg) and did not regain goal pressure until leakage was stopped. With leakage from 12 mm port site, the PBI pressure decreased to 9-11 mmHg, whereas the SI1 and SI2 lost insufflation pressures completely. The PBI maintained pressure of >11 mmHg during continuous suction while the SI1 and SI2 lost pressure entirely, and actually showed negative pressure from air suction into the rigid box system. When evaluated statistically with the mixed model repeated measures ANOVA, the SI1 and SI2 performed similarly while the PBI maintained increased pressure.

conclusionsIn the experimental rigid box system, the PBI more successfully maintained pressure in response to leakage and suction than SIs.

Indexed as

Analysis of VarianceEquipment DesignEquipment FailureInsufflationManometryPneumoperitoneum, ArtificialPressure

Identifiers

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked