Trial report in Clinical orthopaedics and related research, 2012. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 33 papers, 2 of them syntheses that pooled it.

0numbers the graph read from it
0cells of the map it votes in
33citing papers in PubMed, 2 pooled it
12.0field-weighted citation impact, top 1% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

33 citing papers in PubMed, 2 syntheses or guidelines pooled it, 103 citations in OpenAlex.

  1. Pooled it
  2. Pooled it
  3. Trial
  4. Trial
  5. Article
  6. Article
  7. Article
  8. Article
  9. Article
  10. Article
  11. Article
  12. Article
  13. Article
  14. Article
  15. Article
  16. Article
  17. Article
  18. Article
  19. Article
  20. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 2 institutions in 1 country.

Per-Erik JohansonDepartment of Orthopaedics, Institution of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, 413 45 Gothenburg, Sweden. per-erik.johanson@vgregion.se
Georgios Digas
Peter Herberts
Jonas Thanner
Johan Kärrholm
University of Gothenburg · SESahlgrenska University Hospital · SE

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundPolyethylene (PE) wear particles are believed to cause aseptic loosening and thereby impair function in hip arthroplasty. Highly crosslinked polyethylene (XLPE) has low short- and medium-term wear rates. However, the long-term wear characteristics are unknown and it is unclear whether reduced wear particle burden improves function and survival of cemented hip arthroplasty. QUESTIONS/PURPOSES: We asked whether XLPE wear rates remain low up to 10 years and whether this leads to improved implant fixation, periprosthetic bone quality, and clinical function compared to conventional PE.

methodsWe randomized 60 patients (61 hips) to receive either PE or XLPE cemented cups combined with a cemented stem. At 10 years postoperatively, 51 patients (52 hips) were evaluated for polyethylene wear and component migration estimation by radiostereometry, for radiolucent lines, bone densitometry, and Harris hip and pain scores. Revisions were recorded.

resultsXLPE cups had a lower mean three-dimensional wear rate between 2 and 10 years compared to conventional PE hips: 0.005 mm/year versus 0.056 mm/year. We found no differences in cup migration, bone mineral density, radiolucencies, functional scores, and revision rate. There was a trend toward improved stem fixation in the XLPE group. The overall stem failure rate was comparably high, without influencing wear rate in XLPE hips.

conclusionsXLPE displayed a low wear rate up to 10 years when used in cemented THA, but we found no clear benefits in any other parameters. Further research is needed to determine whether cemented THA designs with XLPE are less prone to stem loosening. LEVEL OF EVIDENCE: Level I, therapeutic study. See the Instructions for Authors for a complete description of levels of evidence.

Indexed as

Biocompatible MaterialsPolyethyleneAdultAgedArthroplasty, Replacement, HipBone CementsFemaleFollow-Up StudiesHip JointHip ProsthesisHumansMaleMiddle AgedRadiographyBiocompatible MaterialsBone CementsPolyethylene

Identifiers

PMID22669546
PMCPMC3462851
OpenAlexW2057960802