Evidence map›Paper›PMID 22538673›Full record

ArticleSurgical endoscopy2012

Development of novel electrospun absorbable polycaprolactone (PCL) scaffolds for hernia repair applications.

Gregory C Ebersole, Evan G Buettmann, Matthew R MacEwan, Michael E Tang, Margaret M Frisella, Brent D Matthews, Corey R Deeken

Abstract read
PubMed Publisher
In one paragraph

Article in Surgical endoscopy, 2012. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 15 papers.

0numbers the graph read from it
0cells of the map it votes in
15citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

15 citing papers in PubMed.

  1. Article
  2. Article
  3. Review
  4. Article
  5. Article
  6. Frontiers in bioengineering and biotechnology · 2020
    Article
  7. Article
  8. Article
  9. Article
  10. Article
  11. Review
  12. Article
  13. Article
  14. Article
  15. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Gregory C EbersoleSection of Minimally Invasive Surgery, Department of Surgery, Washington University School of Medicine, 660 South Euclid Avenue, Campus Box 8109, St. Louis, MO 63110, USA.
Evan G Buettmann
Matthew R MacEwan
Michael E Tang
Margaret M Frisella
Brent D Matthews
Corey R Deeken

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

introductionPermanent/nonresorbable hernia repair materials rely on profibrotic wound healing, and repair sites are commonly composed of disorganized tissue with inferior mechanical strength and risk of reherniation. Resorbable electrospun scaffolds represent a novel class of biomaterials, which may provide a unique platform for the design of advanced soft tissue repair materials. These materials are simple, inexpensive, nonwoven materials composed of polymer fibers that readily mimic the natural extracellular matrix. The primary goal of the present study was to evaluate the physiomechanical properties of novel electrospun scaffolds to determine their suitability for hernia repair. Based on previous experimentation, scaffolds possessing ≥ 20 N suture retention strength, ≥ 20 N tear resistance, and ≥ 50 N/cm tensile strength are appropriate for hernia repair.

methodsSix novel electrospun scaffolds were fabricated by varying combinations of polymer concentration (10-12 %) and flow rate (3.5-10 mL/h). Briefly, poly(ε-caprolactone) (PCL) was dissolved in a solvent mixture and electrospun onto a planar metal collector, yielding sheets with randomly oriented fibers. Physiomechanical properties were evaluated through scanning electron microscopy, laser micrometry, and mechanical testing.

resultsScanning electron micrographs demonstrated fiber diameters ranging from 1.0 ± 0.1 μm (10 % PCL, 3.5 mL/h) to 1.5 ± 0.2 μm (12 % PCL, 4 mL/h). Laser micrometry demonstrated thicknesses ranging from 0.72 ± 0.07 mm (12 % PCL, 10 mL/h) to 0.91 ± 0.05 mm (10 % PCL, 3.5 mL/h). Mechanical testing identified two scaffolds possessing suture retention strengths ≥ 20 N (12 % PCL, 10 mL/h and 12 % PCL, 6 mL/h), and no scaffolds possessing tear resistance values ≥ 20 N (range, 4.7 ± 0.9 N to 10.6 ± 1.8 N). Tensile strengths ranged from 35.27 ± 2.08 N/cm (10 % PCL, 3.5 mL/h) to 81.76 ± 15.85 N/cm (12 % PCL, 4 mL/h), with three scaffolds possessing strengths ≥ 50 N/cm (12 % PCL, 10 mL/h; 12 % PCL, 6 mL/h; 12 % PCL, 4 mL/h).

conclusionsTwo electrospun scaffolds (12 % PCL, 10 mL/h and 12 % PCL, 6 mL/h) possessed suture retention and tensile strengths appropriate for hernia repair, justifying evaluation in a large animal model. Additional studies examining advanced methods of fabrication may further improve the unique properties of these scaffolds, propelling them into applications in a variety of clinical settings.

Indexed as

Absorbable ImplantsBiocompatible MaterialsMaterials TestingPolyestersSurgical MeshTissue ScaffoldsHerniorrhaphyHumansStress, MechanicalSuturesTensile StrengthTissue EngineeringBiocompatible MaterialspolycaprolactonePolyesters

Identifiers

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked