Development of a highly efficient implanted thermal ablation device: in vivo experiment in rat liver.

H Matsui, M Hamuro, K Nakamura, H Kayahara, K Murano, Y Kotsuka, Y Miki

Abstract read
In one paragraph

Article in The British journal of radiology, 2012. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

H MatsuiDepartment of Radiology, Osaka City University Graduate School of Medicine, Osaka, Japan. chu_3@i.707.to
M Hamuro
K Nakamura
H Kayahara
K Murano
Y Kotsuka
Y Miki

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectivesTo evaluate an implanted thermal ablation device that can be heated with high efficiency using a resonant circuit as the implant.

methods16 rats were used. The implants, adjusted at a resonance frequency of 4 MHz, were fixed on the surface of the liver of rats under laparotomy. In 14 of 16 rats, an alternating magnetic field (AMF) was applied for 6 min with an output of 300 W from outside the body using a ferrite core applicator. The implant temperature during AMF exposure was measured. The 14 rats were divided into 5 groups, depending on time from AMF application until they were sacrificed (1 h, 1 day, 3 days, 7 days and 1 month after application). Two rats not exposed to AMF were used as controls. Livers were removed and evaluated; the cross-sectional area and width of the ablated region were measured.

resultsDuring AMF exposure, the implant temperature rose to 127.8±39.3 °C (mean±standard deviation). The cross-sectional area of the ablated region was largest after 1 day and tended to decrease with time. The widths of the ablated region were 4.87±0.22 mm, 4.15±0.36 mm, 3.67±0.58 mm and 3.24±0.16 mm in the 1 day, 3 day, 7 day and 1 month groups, respectively. No significant differences (p<0.05) were seen in either cross-sectional area or width of the ablated region.

conclusionSufficient heat for ablation was obtained in vivo using a newly developed implanted thermal ablation device. This device may be a new option for thermal ablation therapy.

Indexed as

AnimalsElectrosurgeryEquipment DesignEquipment Failure AnalysisHepatectomyHyperthermia, InducedLiverMagneticsMalePilot ProjectsRats