SynthesisBMJ (Clinical research ed.)2011
Comparative assessment of implantable hip devices with different bearing surfaces: systematic appraisal of evidence.
Synthesis in BMJ (Clinical research ed.), 2011. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 18 papers, 3 of them syntheses that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
18 citing papers in PubMed, 3 syntheses or guidelines pooled it, 53 citations in OpenAlex.
- Appraisal of evidence base for introduction of new implants in hip and knee replacement: a systematic review of five widely used device technologies.BMJ (Clinical research ed.) · 2014Pooled it
- Total hip replacement for the treatment of end stage arthritis of the hip: a systematic review and meta-analysis.PloS one · 2014Pooled it
- Total disc replacement for chronic back pain in the presence of disc degeneration.The Cochrane database of systematic reviews · 2012Pooled it
- Advancing the Real-World Evidence for Medical Devices through Coordinated Registry Networks.BMJ surgery, interventions, & health technologies · 2022Article
- Post-marketing surveillance on safety and efficacy of posterior spinal correction and fusion with the CD Horizon Solera instrumentation for adolescent idiopathic scoliosis. A retrospective cohort study.North American Spine Society journal · 2021Article
- Large Diameter Head in Primary Total Hip Arthroplasty: A Systematic Review.Indian journal of orthopaedics · 2020Review
- Cemented Exeter total hip arthroplasty with a 32 mm head on highly crosslinked polyethylene: Does age influence functional outcome, satisfaction, activity, stem migration, and periprosthetic bone mineral density?Bone & joint research · 2019Article
- Total Hip Replacement Revision in a Single Brand Small Cementless Stem - Our Experience after the Findings of the National Joint Registry.The archives of bone and joint surgery · 2018Article
- Bearing surfaces in primary total hip arthroplasty.EFORT open reviews · 2018Article
- The role of national registries in improving patient safety for hip and knee replacements.BMC musculoskeletal disorders · 2017Article
- Advances in hip arthroplasty surgery: what is justified?EFORT open reviews · 2017Article
- Have cementless and resurfacing components improved the medium-term results of hip replacement for patients under 60 years of age?Acta orthopaedica · 2015Article
- No functional benefit of larger femoral heads and alternative bearings at 6 months following primary hip replacement.Acta orthopaedica · 2015Article
- Article
- Comparative effectiveness of ceramic-on-ceramic implants in stemmed hip replacement: a multinational study of six national and regional registries.The Journal of bone and joint surgery. American volume · 2014Article
- Test-retest reliability of Common Mental Disorders Questionnaire (CMDQ) in patients with total hip replacement (THR).BMC psychology · 2014Article
- Modular sleeves with ceramic heads in isolated acetabular cup revision in younger patients-laboratory and experimental analysis of suitability and clinical outcomes.International orthopaedics · 2013Article
- Results of 3,668 primary total hip replacements.Acta orthopaedica · 2012Article
Corrections and comments
- Commented on by
Authors and funding
6 authors at 3 institutions in 2 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectiveTo determine comparative safety and effectiveness of combinations of bearing surfaces of hip implants.
designSystematic review of clinical trials, observational studies, and registries. DATA SOURCES: Medline, Embase, Cochrane Controlled Trials Register, reference lists of articles, annual reports of major registries, summaries of safety and effectiveness for pre-market application and mandated post-market studies at the United States Food and Drug Administration. STUDY SELECTION: Criteria for inclusion were comparative studies in adults reporting information for various combinations of bearings (such as metal on metal and ceramic on ceramic). Data search, abstraction, and analyses were independently performed and confirmed by at least two authors. Qualitative data syntheses were performed.
resultsThere were 3139 patients and 3404 hips enrolled in 18 comparative studies and over 830 000 operations in national registries. The mean age range in the trials was 42-71, and 26-88% were women. Disease specific functional outcomes and general quality of life scores were no different or they favoured patients receiving metal on polyethylene rather than metal on metal in the trials. While one clinical study reported fewer dislocations associated with metal on metal implants, in the three largest national registries there was evidence of higher rates of implant revision associated with metal on metal implants compared with metal on polyethylene. One trial reported fewer revisions with ceramic on ceramic compared with metal on polyethylene implants, but data from national registries did not support this finding.
conclusionsThere is limited evidence regarding comparative effectiveness of various hip implant bearings. Results do not indicate any advantage for metal on metal or ceramic on ceramic implants compared with traditional metal on polyethylene or ceramic on polyethylene bearings.
Indexed as