Dose accuracy of the ClikSTAR, NovoPen 4, and Luxura insulin pens: results of laboratory and field studies.
Article in Journal of diabetes science and technology, 2011. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
- Half-Unit Insulin Pens: Disease Management in Patients With Diabetes Who Are Sensitive to Insulin.Journal of diabetes science and technology · 2017Review
- Dialing torque preferences of people with diabetes when using insulin pens: a pilot study.Diabetes therapy : research, treatment and education of diabetes and related disorders · 2015Article
- Analysis: linking laboratory data to human factors and inclusion of persons with disabilities in diabetes technology research.Journal of diabetes science and technology · 2011Article
Corrections and comments
- Commented on by
Authors and funding
3 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundA high dosing accuracy is needed to maintain normal glycemia in patients with diabetes. This study investigated the dose accuracy of the commonly used reusable insulin pens ClikSTAR®, NovoPen® 4, and Luxura®.
methodsPens were tested in a laboratory setting by one trained technician who delivered four doses of 30 U from each of 15 pens per pen model (a total of 60 doses from each pen model). Pens were also tested in a simulated clinical setting by 48 people with diabetes. Each participant delivered 27 doses: three doses of 30 U from each of three pens per pen model. Overall, the technician delivered 180 doses and the participants 1296 doses.
resultsAll pens met the tolerance limits defined by the German edition of the International Standardization Organization (ISO) 11608-1:2000 standard [30 ± 1.5 U (28.5-31.5 U)]. All doses were delivered within the limits proposed by the ISO, except for two doses with Luxura in the clinical setting. In laboratory testing, the mean dose delivered by ClikSTAR (29.69 U) or Luxura (29.89 U) was less than the expected 30 U and significantly less than the mean dose delivered by NovoPen 4 (30.04 U; p < .001 for both comparisons). Similar results were observed in the simulated clinical setting. NovoPen 4 had the greatest variance in laboratory testing but the least in the simulated clinical setting.
conclusionsThis study demonstrates comparable dose accuracy and variability of the ClikSTAR, Luxura, and NovoPen 4 insulin pens. The slight differences in mean doses between pens are unlikely to be clinically significant.