ArticleJournal of diabetes science and technology2011
Performance evaluation and labeling comprehension of a new blood glucose monitoring system with integrated information management.
Article in Journal of diabetes science and technology, 2011. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed, 4 citations in OpenAlex.
- Analysis of the evaluation of a new glucose meter with integrated self-management software and USB connectivity.Journal of diabetes science and technology · 2011Article
Corrections and comments
- Commented on by
Authors and funding
8 authors at 2 institutions in 2 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundThis study evaluated performance and product labeling of CONTOUR® USB, a new blood glucose monitoring system (BGMS) with integrated diabetes management software and a universal serial bus (USB) port, in the hands of untrained lay users and health care professionals (HCPs).
methodSubjects and HCPs tested subject's finger stick capillary blood in parallel using CONTOUR USB meters; deep finger stick blood was tested on a Yellow Springs Instruments (YSI) glucose analyzer for reference. Duplicate results by both subjects and HCPs were obtained to assess system precision. System accuracy was assessed according to International Organization for Standardization (ISO) 15197:2003 guidelines [within ±15 mg/dl of mean YSI results (samples <75 mg/dl) and ±20% (samples ≥75 mg/dl)]. Clinical accuracy was determined by Parkes error grid analysis. Subject labeling comprehension was assessed by HCP ratings of subject proficiency. Key system features and ease-of-use were evaluated by subject questionnaires.
resultsAll subjects who completed the study (N = 74) successfully performed blood glucose measurements, connected the meter to a laptop computer, and used key features of the system. The system was accurate; 98.6% (146/148) of subject results and 96.6% (143/148) of HCP results exceeded ISO 15197:2003 criteria. All subject and HCP results were clinically accurate (97.3%; zone A) or associated with benign errors (2.7%; zone B). The majority of subjects rated features of the BGMS as "very good" or "excellent."
conclusionsCONTOUR USB exceeded ISO 15197:2003 system performance criteria in the hands of untrained lay users. Subjects understood the product labeling, found the system easy to use, and successfully performed blood glucose testing.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.