Evidence map›Paper›PMID 22003351›Full record

ReviewJournal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine2011

Obstructive sleep apnea devices for out-of-center (OOC) testing: technology evaluation.

Nancy A Collop, Sharon L Tracy, Vishesh Kapur, Reena Mehra, David Kuhlmann, Sam A Fleishman, Joseph M Ojile

Registry-linked trialOpen access · bronzeAbstract readReview
In one paragraph

Review in Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine, 2011. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT04752462 (Effect of Post-Discharge Telemedicine Motivational Enhancement Intervention in Heart Failure and Obstructive Sleep Apnea on Adherence and Health Outcomes), which is not on this map. Cited by 189 papers, 5 of them syntheses that pooled it.

0numbers the graph read from it
0cells of the map it votes in
189citing papers in PubMed, 5 pooled it
9.9field-weighted citation impact, top 1% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

NCT04752462 nacompletednot on this mapstarted 2021, after this paper: background citation

Effect of Post-Discharge Telemedicine Motivational Enhancement Intervention in Heart Failure and Obstructive Sleep Apnea on Adherence and Health Outcomes

TypeinterventionalSponsorThe Cleveland ClinicRan2021 to 2025Enrolled75ConditionsAcute Decompensated Heart Failure, Obstructive Sleep Apnea, Motivational Enhancement, PAP AdherenceArmsTelemedicine Intensive Motivational Enhancement
3 · Its place in the literature

Who cites it

189 citing papers in PubMed, 5 syntheses or guidelines pooled it, 420 citations in OpenAlex.

  1. Pooled it
  2. Guideline
  3. Clinical Practice Guideline for Diagnostic Testing for Adult Obstructive Sleep Apnea: An American Academy of Sleep Medicine Clinical Practice Guideline.Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine · 2017
    Guideline
  4. Acoustic Analysis of Snoring in the Diagnosis of Obstructive Sleep Apnea Syndrome: A Call for More Rigorous Studies.Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine · 2015
    Pooled it
  5. Guideline
  6. Trial
  7. Long-term night-to-night variability of sleep-disordered breathing using a radar-based home sleep apnea test: a prospective cohort study.Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine · 2024
    Trial
  8. Validation of the Withings Sleep Analyzer, an under-the-mattress device for the detection of moderate-severe sleep apnea syndrome.Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine · 2021
    Trial
  9. Trial
  10. Trial
  11. Trial
  12. A Comprehensive Evaluation of a Two-Channel Portable Monitor to "Rule in" Obstructive Sleep Apnea.Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine · 2015
    Trial
  13. Home-based diagnosis of obstructive sleep apnea in an urban population.Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine · 2014
    Trial
  14. Diagnosis of obstructive sleep apnea using pulse oximeter derived photoplethysmographic signals.Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine · 2014
    Trial
  15. Objective and subjective adherence to oral appliance therapy in adults with obstructive sleep apnea: pilot data from a one-year longitudinal study.Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine · 2026
    Article
  16. From classic to cutting-edge: technological approaches to respiratory physiological signals in assessing sleep-disordered breathing.Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine · 2026
    Review
  17. Article
  18. Article
  19. Article
  20. Sleep Medicine-What's in a Name?Journal of sleep research · 2025
    Review

129 more citing papers are in PubMed but not listed here.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 7 institutions in 1 country.

Nancy A CollopEmory Sleep Center, Atlanta, GA, USA.
Sharon L Tracy
Vishesh Kapur
Reena Mehra
David Kuhlmann
Sam A Fleishman
Joseph M Ojile
American Academy of Sleep Medicine · USCape Fear Valley Medical Center · USCase Western Reserve University · USClayton Sleep Institute · USEmory Healthcare · USRegional Health · USUniversity of Washington · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Guidance is needed to help clinicians decide which out-of-center (OOC) testing devices are appropriate for diagnosing obstructive sleep apnea (OSA). A new classification system that details the type of signals measured by these devices is presented. This proposed system categorizes OOC devices based on measurements of Sleep, Cardiovascular, Oximetry, Position, Effort, and Respiratory (SCOPER) parameters.Criteria for evaluating the devices are also presented, which were generated from chosen pre-test and post-test probabilities. These criteria state that in patients with a high pretest probability of having OSA, the OOC testing device has a positive likelihood ratio (LR+) of 5 or greater coinciding with an in-lab-polysomnography (PSG)-generated apnea hypopnea index (AHI) ≥ 5, and an adequate sensitivity (at least 0.825).Since oximetry is a mandatory signal for scoring AHI using PSG, devices that do not incorporate oximetry were excluded. English peer-reviewed literature on FDA-approved devices utilizing more than 1 signal was reviewed according to the above criteria for 6 questions. These questions specifically addressed the adequacy of different respiratory and effort sensors and combinations thereof to diagnose OSA. In summary, the literature is currently inadequate to state with confidence that a thermistor alone without any effort sensor is adequate to diagnose OSA; if a thermal sensing device is used as the only measure of respiration, 2 effort belts are required as part of the montage and piezoelectric belts are acceptable in this context; nasal pressure can be an adequate measurement of respiration with no effort measure with the caveat that this may be device specific; nasal pressure may be used in combination with either 2 piezoelectric or respiratory inductance plethysmographic (RIP) belts (but not 1 piezoelectric belt); and there is insufficient evidence to state that both nasal pressure and thermistor are required to adequately diagnose OSA. With respect to alternative devices for diagnosing OSA, the data indicate that peripheral arterial tonometry (PAT) devices are adequate for the proposed use; the device based on cardiac signals shows promise, but more study is required as it has not been tested in the home setting; for the device based on end-tidal CO(2) (ETCO(2)), it appears to be adequate for a hospital population; and for devices utilizing acoustic signals, the data are insufficient to determine whether the use of acoustic signals with other signals as a substitute for airflow is adequate to diagnose OSA.Standardized research is needed on OOC devices that report LR+ at the appropriate AHI (≥ 5) and scored according to the recommended definitions, while using appropriate research reporting and methodology to minimize bias.

Indexed as

Equipment DesignHumansMonitoring, AmbulatoryOximetryPlethysmography, ImpedancePredictive Value of TestsRespiratory Function TestsSleep Apnea, ObstructiveTransducers, PressureAHIObstructive sleep apneaout-of-center testingSCOPERtesting devices

Identifiers

PMID22003351
PMCPMC3190855
OpenAlexW2093372772

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials