ArticleDiabetes technology & therapeutics2011

Glucose response of near-infrared alginate-based microsphere sensors under dynamic reversible conditions.

Ayesha Chaudhary, Harri Harma, Pekka Hanninen, Michael J McShane, Rohit Srivastava

Abstract read
In one paragraph

Article in Diabetes technology & therapeutics, 2011. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed.

  1. Alginate NiFeGels (Basel, Switzerland) · 2021
    Article
  2. Article
  3. Article
  4. Gel encapsulation of glucose nanosensors for prolonged in vivo lifetime.Journal of diabetes science and technology · 2013
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Ayesha ChaudharyDepartment of Biosciences and Bioengineering, IIT Bombay, Powai, Mumbai, India.
Harri Harma
Pekka Hanninen
Michael J McShane
Rohit Srivastava

Funding

Fluorescent Glucose Sensors from Polyion MicroshellsR01EB000739 · NIBIB · TEXAS ENGINEERING EXPERIMENT STATION · PI MCSHANE, MIKE · 2002 to 2006
$1.3M
NIBIB NIH HHS R01 EB00739-5
6 · The paper itself

Abstract

backgroundMinimally invasive optical glucose biosensors with increased functional longevity form one of the most promising techniques for continuous glucose monitoring, because of their long-term stability, reversibility, repeatability, specificity, and high sensitivity. They are based on the principle of competitive binding and fluorescence resonance energy transfer. Moving to the near-infrared region of the spectrum has the potential to improve signal throughput for implanted sensors, but requires a change in dye chemistry that could alter response sensitivity, range, and toxicity profiles.

methodsThe near-infrared dissolved-core alginate microsphere sensors were fabricated by emulsion followed by surface coating by layer-by-layer self-assembly. The particles were characterized for sensor stability, sensor response, and reversibility in deionized water and simulated interstitial fluid. The sensor response to step changes in bulk glucose concentrations was also evaluated under dynamic conditions using a microflow cell unit. Finally, in vitro cytotoxicity assays were performed with L929 mouse fibroblast cell lines to demonstrate preliminary biocompatibility of the sensors.

resultsThe glucose sensitivity under controlled and dynamic conditions was observed to be 0.86%/mM glucose with an analytical response range of 0-30 mM glucose, covering both the physiological and pathophysiological range. The sensor demonstrated a repeatable, reversible, and reproducible response, with a maximum response time of 120 s. In vitro cytotoxicity assays revealed nearly 95% viability of cells, thereby suggesting that the alginate microsphere sensor system does not exhibit cytotoxicity.

conclusionsThe incorporation of near-infrared dyes shows promise in improving sensor response because of their high sensitivity and improved tissue penetration of infrared light. The sensitivity for the sensors was approximately 1.5 times greater than that observed for visible dye sensors, and the new dye chemistry did not significantly alter the biocompatibility of the materials. These findings provide additional support for the potential application of alginate microspheres and similar systems such as "smart-tattoo" glucose sensors.

Indexed as

MicrospheresAlginatesAnimalsBiosensing TechniquesBlood GlucoseCell LineDiabetes MellitusFluorescent DyesMiceReproducibility of ResultsSpectrometry, FluorescenceAlginatesBlood GlucoseFluorescent Dyes

Identifiers

PMID21568749
PMCPMC3133682

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.