Evidence map›Paper›PMID 21153399›Full record

ArticleIntensive care medicine2011

Arterial pressure-based cardiac output monitoring: a multicenter validation of the third-generation software in septic patients.

Daniel De Backer, Gernot Marx, Andrew Tan, Christopher Junker, Marc Van Nuffelen, Lars Hüter, Willy Ching, Frédéric Michard, Jean-Louis Vincent

Open access · hybridAbstract readMulticenter StudyValidation Study
In one paragraph

Article in Intensive care medicine, 2011. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 53 papers, 2 of them syntheses that pooled it.

0numbers the graph read from it
0cells of the map it votes in
53citing papers in PubMed, 2 pooled it
14.8field-weighted citation impact, top 1% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

53 citing papers in PubMed, 2 syntheses or guidelines pooled it, 137 citations in OpenAlex.

  1. Pooled it
  2. Guideline
  3. Trial
  4. Trial
  5. Trial
  6. Trial
  7. Trial
  8. Observational
  9. Article
  10. Article
  11. Article
  12. Monitoring cardiac output.Intensive care medicine · 2024
    Article
  13. Observational
  14. Article
  15. Observational
  16. Article
  17. Review
  18. Article
  19. Review
  20. Article
4 · The record

Corrections and comments

5 · Who and what money

Authors and funding

9 authors at 6 institutions in 4 countries.

Daniel De BackerDepartment of Intensive Care, Erasme University Hospital, Université Libre de Bruxelles, Route de Lennik 808, 1070 Brussels, Belgium. ddebacke@ulb.ac.be
Gernot Marx
Andrew Tan
Christopher Junker
Marc Van Nuffelen
Lars Hüter
Willy Ching
Frédéric Michard
Jean-Louis Vincent
Université Libre de Bruxelles · BEQueen's Medical Center · USEdwards Lifesciences (Switzerland) · CHGeorge Washington University · USJena University Hospital · DEUniversitätsklinikum Aachen · DE

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeSecond-generation FloTrac software has been shown to reliably measure cardiac output (CO) in cardiac surgical patients. However, concerns have been raised regarding its accuracy in vasoplegic states. The aim of the present multicenter study was to investigate the accuracy of the third-generation software in patients with sepsis, particularly when total systemic vascular resistance (TSVR) is low.

methodsFifty-eight septic patients were included in this prospective observational study in four university-affiliated ICUs. Reference CO was measured by bolus pulmonary thermodilution (iCO) using 3-5 cold saline boluses. Simultaneously, CO was computed from the arterial pressure curve recorded on a computer using the second-generation (CO(G2)) and third-generation (CO(G3)) FloTrac software. CO was also measured by semi-continuous pulmonary thermodilution (CCO).

resultsA total of 401 simultaneous measurements of iCO, CO(G2), CO(G3), and CCO were recorded. The mean (95%CI) biases between CO(G2) and iCO, CO(G3) and iCO, and CCO and iCO were -10 (-15 to -5)% [-0.8 (-1.1 to -0.4) L/min], 0 (-4 to 4)% [0 (-0.3 to 0.3) L/min], and 9 (6-13)% [0.7 (0.5-1.0) L/min], respectively. The percentage errors were 29 (20-37)% for CO(G2), 30 (24-37)% for CO(G3), and 28 (22-34)% for CCO. The difference between iCO and CO(G2) was significantly correlated with TSVR (r(2) = 0.37, p < 0.0001). A very weak (r(2) = 0.05) relationship was also observed for the difference between iCO and CO(G3).

conclusionsIn patients with sepsis, the third-generation FloTrac software is more accurate, as precise, and less influenced by TSVR than the second-generation software.

Indexed as

Catheterization, Swan-GanzSoftwareAgedBlood PressureCardiac OutputFemaleHumansIntensive Care UnitsMaleMiddle AgedMonitoring, PhysiologicSepsisVasoplegia

Identifiers

PMID21153399
PMCPMC3028067
OpenAlexW2084638854

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked