Evidence map›Paper›PMID 21127827›Full record

GuidelineDer Anaesthesist2010

[Prevention of infections under anesthetic breathing with breathing filters: concerted recommendations of the Deutsche Gesellschaft für Krankenhaushygiene e.V. (DGKH) and the Deutsche Gesellschaft für Anästhesiologie und Intensivmedizin e.V. (DGAI)].

R Kranabetter, A Kramer, J Rathgeber, K Züchner, O Assadian, G Daeschlein, N-O Hübner, E Dietlein, M Exner, M Gründling and 10 more

Open access · bronzeAbstract readEnglish AbstractGuideline
In one paragraph

Guideline in Der Anaesthesist, 2010. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
0.2field-weighted citation impact, top 39% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 9 citations in OpenAlex.

  1. Review
  2. [Infection Prevention in Anesthesiology during the SARS-CoV 2-pandemic].Anasthesiologie, Intensivmedizin, Notfallmedizin, Schmerztherapie : AINS · 2020
    Article
  3. Review
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

20 authors at 1 institution in 1 country.

R Kranabetterhr.kranabetter@o2online.de
A Kramer
J Rathgeber
K Züchner
O Assadian
G Daeschlein
N-O Hübner
E Dietlein
M Exner
M Gründling
C Lehmann
M Wendt
B M Graf
D Holst
L Jatzwauk
B Puhlmann
T Welte
A R Wilkes
Deutschen Gesellschaft für Krankenhaushygiene
Deutschen Gesellschaft für Anästhesiologie und Intensivmedizin
Georg Simon Ohm University of Applied Sciences Nuremberg · DE

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

An interdisciplinary working group from the German Society of Hospital Hygiene (DGKH) and the German Society for Anesthesiology and Intensive Care (DGAI) worked out the following recommendations for infection prevention during anesthesia by using breathing system filters (BSF). The BSF shall be changed after each patient. The filter retention efficiency for airborne particles is recommended to be >99% (II). The retention performance of BSF for liquids is recommended to be at pressures of at least 60 hPa (=60 mbar) or 20 hPa above the selected maximum ventilation pressure in the anesthetic system.The anesthesia breathing system may be used for a period of up to 7 days provided that the functional requirements of the system remain unchanged and the manufacturer states this in the instructions for use. The breathing system and the manual ventilation bag are changed immediately after the respective anesthesia if the following situation has occurred or it is suspected to have occurred: Notifiable infectious disease involving the risk of transmission via the breathing system and the manual bag, e.g. tuberculosis, acute viral hepatitis, measles, influenza virus, infection and/or colonization with a multi-resistant pathogen or upper or lower respiratory tract infections. In case of visible contamination e.g. by blood or in case of defect, it is required that the BSF and also the anesthesia breathing system is changed and the breathing gas conducting parts of the anesthesia ventilator are hygienically reprocessed.Observing of the appropriate hand disinfection is very important. All surfaces of the anesthesia equipment exposed to hand contact must be disinfected after each case.

Indexed as

Anesthesia, InhalationAnesthetics, InhalationCross InfectionFiltrationHumansHygieneAnesthetics, Inhalation

Identifiers

PMID21127827
PMCPMC7080057
OpenAlexW1510645953

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.