ArticleActa orthopaedica2010
Knee flexion influences periprosthetic BMD measurement in the tibia. Suggestions for a reproducible clinical scan protocol.
Article in Acta orthopaedica, 2010. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
8 citing papers in PubMed, 27 citations in OpenAlex.
- Comparison of cementless twin-peg, cemented twin-peg and cemented single-peg femoral component migration after medial unicompartmental knee replacement: a 5-year randomized RSA study.Archives of orthopaedic and trauma surgery · 2023Trial
- Superior fixation and less periprosthetic stress-shielding of tibial components with a finned stem versus an I-beam block stem: a randomized RSA and DXA study with minimum 5 years' follow-up.Acta orthopaedica · 2019Trial
- Mobile- vs. fixed-bearing total knee replacement.Acta orthopaedica · 2015Trial
- Changes in Periprosthetic Bone Mineral Density Following Arthroplasty: An In-Depth Review and Current Perspectives.Current osteoporosis reports · 2025Review
- Periprosthetic fractures after medial unicompartmental knee arthroplasty: a narrative review.Archives of orthopaedic and trauma surgery · 2022Review
- [Progress of change in bone mineral density after knee arthroplasty].Zhongguo xiu fu chong jian wai ke za zhi = Zhongguo xiufu chongjian waike zazhi = Chinese journal of reparative and reconstructive surgery · 2021Article
- Multivariable logistic and linear regression models for identification of clinically useful biomarkers for osteoarthritis.Scientific reports · 2020Article
- Fatigue fracture of tibial arthroplasty implant masked by contralateral knee arthritis.International journal of surgery case reports · 2013Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors at 4 institutions in 2 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
background and purposeThe quality and quantity of bone is important for the success of joint prostheses and may be monitored by dual energy X-ray absorptiometry (DXA). Available protocols suggest that the knee should be positioned in full extension. This is not possible for most patients in the first days after surgery; however, deficits in extension normalize with rehabilitation. Individual knee flexion between the baseline and follow-up investigations may therefore be different. We investigated the sensitivity of bone mineral density (BMD) measurements to knee flexion in a phantom study and in patients. We suggest a protocol for clinical use.
methods2 phantom tibial bones with tibia components were secured in a clamp and BMD measurements were repeated 5 times at every 5 degrees change in flexion from 0 degrees to 20 degrees. For clinical use, a soft foam positioner was produced, in which the lower leg could be placed in neutral rotation and with the knee in approximately 25 degrees of flexion. The clinical repeatability was tested with double examinations in 38 patients. We investigated 3 regions of interest (ROIs) below the tibial plateau.
resultsIn the phantom study, just 5 degrees of flexion was found to change the measured mean BMD. The reproducibility of clinical measurements (coefficient of variation) in the 3 ROIs assessed ranged from 1.8% to 3.7% for the anteroposterior scans, and from 3.4% to 6.2% for the lateral scans.
interpretationKnee flexion does affect the measured periprosthetic tibial BMD, and knee flexion should be the same at all clinical follow-ups. The protocol and soft foam positioner that we suggest permit precise and reliable assessment of BMD in the proximal tibia and they can be used in clinical work.
Indexed as