ArticleAJNR. American journal of neuroradiology2010
In vivo evaluation of the Phenox CRC mechanical thrombectomy device in a swine model of acute vessel occlusion.
Article in AJNR. American journal of neuroradiology, 2010. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 11 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
11 citing papers in PubMed, 26 citations in OpenAlex.
- Article
- The Protective Effects of Statins Towards Vessel Wall Injury Caused by a Stent Retrieving Mechanical Thrombectomy Device : A Histological Analysis of the Rabbit Carotid Artery Model.Journal of Korean Neurosurgical Society · 2021Article
- Imaging Clot Characteristics in Stroke and its Possible Implication on Treatment.Clinical neuroradiology · 2020Review
- Preclinical Evaluation of the NeVaTM Stent Retriever: Safety and Efficacy in the Swine Thrombectomy Model.Interventional neurology · 2018Article
- Reperfusion therapy in acute ischemic stroke: dawn of a new era?BMC neurology · 2018Review
- What is meant by "TICI"?AJNR. American journal of neuroradiology · 2013Review
- The quest for arterial recanalization in acute ischemic stroke-the past, present and the future.Journal of clinical medicine research · 2013Article
- New stent retriever devices.Current atherosclerosis reports · 2013Review
- Experimental evaluation of immediate recanalization effect and recanalization efficacy of a new thrombus retriever for acute stroke treatment in vivo.AJNR. American journal of neuroradiology · 2013Article
- In vivo evaluation of the first dedicated combined flow-restoration and mechanical thrombectomy device in a swine model of acute vessel occlusion.AJNR. American journal of neuroradiology · 2011Article
- Histopathologic evaluation of arterial wall response to 5 neurovascular mechanical thrombectomy devices in a swine model.AJNR. American journal of neuroradiologyArticle
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
background and purposeMechanical thrombectomy in ischemic stroke is of increasing interest as it is a promising strategy for fast and efficient recanalization. Several thrombectomy devices have been introduced to the armentarium of mechanical thrombectomy. Currently, new devices are under development and are continuously added to the neurointerventional tool box. Each device advocated so far has a different design and mechanical properties in terms of thrombus-device interaction. Therefore, a systematic evaluation under standardized conditions in vivo of these new devices is needed. The purpose of this study was to evaluate the efficiency, thrombus-device interaction, and potential complications of the novel Phenox CRC for distal mechanical thrombectomy in vivo. MATERIALS AND
methodsThe device was evaluated in an established animal model in the swine. Recanalization rate, thromboembolic events, vasospasm, and complications were assessed. Radiopaque thrombi (2 cm length) were used for the visualization of thrombus-device interaction during retrieval. The Phenox CRC (4 mm diameter) was assessed in 15 vessel occlusions. For every occlusion a maximum of 3 retrieval attempts were performed.
resultsComplete recanalization (TICI 3/TIMI 3) was achieved in 86.7% of vessel occlusions. In 66.7% (10/15), the first retrieval attempt was successful, and in 20% (3/15), the second attempt led to complete recanalization of the parent artery. In 2 cases (13.3%) thrombus retrieval was not successful (TICI 0/TIMI 0). In 1 case (6.7%) a minor embolic event occurred in a small side branch. No distal thromboembolic event was observed during the study. Thrombus-device interaction illustrated the entrapment of the thrombus by the microfilaments and the proximal cage of the device. No significant thrombus compression was observed. No vessel perforation, dissection, or fracture of the device occurred.
conclusionsIn this small animal study, the Phenox CRC was a safe and effective device for mechanical thrombectomy. The unique design with a combination of microfilaments and proximal cage reduces thrombus compression with a consequently high recanalization and low complication rate.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials