Trial reportIntensive care medicine2009
Is volume and leak monitoring feasible during nasopharyngeal continuous positive airway pressure in neonates?
Trial report in Intensive care medicine, 2009. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
9 citing papers in PubMed, 22 citations in OpenAlex.
- Non-invasive high-frequency oscillatory ventilation in preterm infants after extubation: a randomized, controlled trial.The Journal of international medical research · 2021Trial
- Work of breathing during HHHFNC and synchronised NIPPV following extubation.European journal of pediatrics · 2019Trial
- Influence of nose and mouth leaks on peripheral oxygen saturation during continuous positive airway pressure in neonates.World journal of pediatrics : WJP · 2013Trial
- A New Computational Method for Detecting Leak Flow and Tidal Volume Waveforms During Spontaneous or Mandatory Breathing Assisted with Nasopharyngeal Ventilation.Sensors (Basel, Switzerland) · 2025Article
- Article
- Respiratory support with heated humidified high flow nasal cannula in preterm infants.Korean journal of pediatrics · 2016Review
- Nasal high-frequency oscillation ventilation in neonates: a survey in five European countries.European journal of pediatrics · 2015Article
- Year in review in Intensive Care Medicine 2009. Part III: mechanical ventilation, acute lung injury and respiratory distress syndrome, pediatrics, ethics, and miscellanea.Intensive care medicine · 2010Review
- Use of prototype bi-nasal prongs for noninvasive ventilation in foals.Journal of veterinary internal medicineArticle
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectiveTo investigate tidal volume (VT) and leak measurements during continuous positive airway pressure (CPAP) in neonates using a commercial ventilatory device equipped with a flow sensor at the Y-piece.
designRandomized cross-over trial.
settingNeonatal intensive care unit level III. PATIENTS: Thirty-two infants, median (range) birth weight 1,435 (710-2,730) g, gestational age 30 (24-38) weeks.
interventionsDuring nasopharyngeal CPAP, leak and VT were measured with and without occlusion of the contralateral nostril using the Leoni ventilator (Heinen & Löwenstein, Germany) and a recently developed algorithm to correct measured VT in the presence of leaks. The measuring range of the Leoni is limited to leaks <90%. MAIN
resultsAnalyzable measurements with leaks <90% could be obtained in 12.5% of the patients with open nostril, and in 65.6% with occluded nostril. Calculated leak flow after nostril occlusion was 23 (3-77) ml min(-1) with closed mouth. Leak flow increased significantly if mouth was opened (548 (0-1,394) ml min(-1), p<0.001), but was probably even higher where leaks exceeded 90%. Mean expiratory volume +/- SD was 5.8 +/- 1.3 ml kg(-1) (corrected VT 5.9 +/-1.2 ml kg(-1)) for leaks <20%, and 3.7 +/-1.4 ml kg(-1) (corrected VT 5.8 +/- 2.2 ml kg(-1)) for leaks between 20 and 69%.
conclusionsLeak and corrected VT could be determined in the presence of leaks of up to 69%, but leaks during CPAP often exceeded the measuring range. Reliable volume and leak monitoring was not possible with the tested equipment during nasopharyngeal CPAP. Advanced equipment is necessary to further investigate the effects of leaks on neonatal CPAP therapy.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials