ArticleIntensive care medicine2009
Evaluation of the user-friendliness of seven new generation intensive care ventilators.
Article in Intensive care medicine, 2009. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 11 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
11 citing papers in PubMed.
- Qualitative and quantitative evaluation of the usability of transport ventilators using eye tracking.Scientific reports · 2026Article
- Ventilator Mode Finder: An Android application to find equivalent of a mode on different ventilators.Journal of research in medical sciences : the official journal of Isfahan University of Medical Sciences · 2021Article
- Comprehensive Evaluation of User Interface for Ventilators Based on Respiratory Therapists' Performance, Workload, and User Experience.Medical science monitor : international medical journal of experimental and clinical research · 2018Article
- Usability Study of the User-Interface of Intensive Care Ventilators Based on User Test and Eye-Tracking Signals.Medical science monitor : international medical journal of experimental and clinical research · 2018Article
- A new global and comprehensive model for ICU ventilator performances evaluation.Annals of intensive care · 2017Article
- The usability of ventilators: a comparative evaluation of use safety and user experience.Critical care (London, England) · 2016Article
- A Critical Review of Mechanical Ventilation Virtual Simulators: Is It Time to Use Them?JMIR medical education · 2016Article
- Healthcare Technology Management (HTM) of mechanical ventilators by clinical engineers.Journal of intensive care · 2014Article
- Year in review in Intensive Care Medicine 2009. Part III: mechanical ventilation, acute lung injury and respiratory distress syndrome, pediatrics, ethics, and miscellanea.Intensive care medicine · 2010Review
- ICU mechanical ventilators, technological advances vs. user friendliness: the right picture is worth a thousand numbers.Intensive care medicine · 2009Article
- The usability of ventilator maintenance user interface: A comparative evaluation of user task performance, workload, and user experience.Science progressArticle
Corrections and comments
- Commented on by
Authors and funding
3 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectiveTo explore the user-friendliness and ergonomics of seven new generation intensive care ventilators.
designProspective task-performing study.
settingIntensive care research laboratory, university hospital.
methodsTen physicians experienced in mechanical ventilation, but without prior knowledge of the ventilators, were asked to perform eight specific tasks [turning the ventilator on; recognizing mode and parameters; recognizing and setting alarms; mode change; finding and activating the pre-oxygenation function; pressure support setting; stand-by; finding and activating non-invasive ventilation (NIV) mode]. The time needed for each task was compared to a reference time (by trained physiotherapist familiar with the devices). A time >180 s was considered a task failure.
resultsFor each of the tests on the ventilators, all physicians' times were significantly higher than the reference time (P < 0.001). A mean of 13 +/- 8 task failures (16%) was observed by the ventilator. The most frequently failed tasks were mode and parameter recognition, starting pressure support and finding the NIV mode. Least often failed tasks were turning on the pre-oxygenation function and alarm recognition and management. Overall, there was substantial heterogeneity between machines, some exhibiting better user-friendliness than others for certain tasks, but no ventilator was clearly better that the others on all points tested.
conclusionsThe present study adds to the available literature outlining the ergonomic shortcomings of mechanical ventilators. These results suggest that closer ties between end-users and manufacturers should be promoted, at an early development phase of these machines, based on the scientific evaluation of the cognitive processes involved by users in the clinical setting.
Indexed as
Identifiers
19639302What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.