Evidence map›Paper›PMID 19559634›Full record

ArticleJournal of magnetic resonance (San Diego, Calif. : 1997)2009

A strip-shield improves the efficiency of a solenoid coil in probes for high-field solid-state NMR of lossy biological samples.

Chin H Wu, Christopher V Grant, Gabriel A Cook, Sang Ho Park, Stanley J Opella

Abstract read
In one paragraph

Article in Journal of magnetic resonance (San Diego, Calif. : 1997), 2009. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 21 papers.

0numbers the graph read from it
0cells of the map it votes in
21citing papers in PubMed
2.5field-weighted citation impact, top 12% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

21 citing papers in PubMed, 31 citations in OpenAlex.

  1. Article
  2. 3D printed sample tubes for solid-state NMR experiments.Journal of magnetic resonance (San Diego, Calif. : 1997) · 2021
    Article
  3. Journal of magnetic resonance (San Diego, Calif. : 1997) · 2020
    Article
  4. Review
  5. Article
  6. Mechanism of dilute-spin-exchange in solid-state NMR.The Journal of chemical physics · 2014
    Article
  7. Article
  8. Article
  9. Article
  10. Heat management strategies for solid-state NMR of functional proteins.Journal of magnetic resonance (San Diego, Calif. : 1997) · 2012
    Article
  11. Article
  12. Article
  13. Review
  14. Article
  15. Article
  16. Article
  17. NMR studies of p7 protein from hepatitis C virus.European biophysics journal : EBJ · 2010
    Article
  18. Probes for high field solid-state NMR of lossy biological samples.Journal of magnetic resonance (San Diego, Calif. : 1997) · 2010
    Review
  19. Article
  20. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 1 institution in 1 country.

Chin H WuDepartment of Chemistry and Biochemistry, 9500 Gilman Drive, University of California, San Diego, La Jolla, CA 92093-0307, USA.
Christopher V Grant
Gabriel A Cook
Sang Ho Park
Stanley J Opella
University of California, San Diego · US

Funding

RESOURCE FOR SOLID STATE NMR OF PROTEINSP41EB002031 · NIBIB · UNIVERSITY OF CALIFORNIA, SAN DIEGO · PI OPELLA, STANLEY J · 2003 to 2018
$16.3M
Molecular Imaging of G-Protein-Coupled Receptors for Drug DevelopmentR01EB005161 · NIBIB · UNIVERSITY OF CALIFORNIA, SAN DIEGO · PI OPELLA, STANLEY J · 2006 to 2014
$6.8M
Structure, Dynamics, and Functions of the Viroporins Vpu from HIV and p7 from HCVR01GM066978 · NIGMS · UNIVERSITY OF CALIFORNIA, SAN DIEGO · PI OPELLA, STANLEY J · 2003 to 2016
$3.5M
Membrane Protein Structures:Phospholipid Bilayers (RMI)R01GM075877 · NIGMS · UNIVERSITY OF CALIFORNIA, SAN DIEGO · PI OPELLA, STANLEY J · 2005 to 2009
$1.4M
NIBIB NIH HHS P41 EB002031NIBIB NIH HHS P41EB002031NIBIB NIH HHS R01 EB005161NIGMS NIH HHS R01 GM066978NIGMS NIH HHS R01 GM075877
6 · The paper itself

Abstract

A strip-shield inserted between a high inductance double-tuned solenoid coil and the glass tube containing the sample improves the efficiency of probes used for high-field solid-state NMR experiments on lossy aqueous samples of proteins and other biopolymers. A strip-shield is a coil liner consisting of thin copper strips layered on a PTFE (polytetrafluoroethylene) insulator. With lossy samples, the shift in tuning frequency is smaller, the reduction in Q, and RF-induced heating are all significantly reduced when the strip-shield is present. The performance of 800MHz (1)H/(15)N and (1)H/(13)C double-resonance probes is demonstrated on aqueous samples of membrane proteins in phospholipid bilayers.

Indexed as

BiopolymersCarbon IsotopesComputer SimulationElectromagnetic FieldsEquipment DesignHydrogenMagnetic Resonance SpectroscopyNitrogen IsotopesPolytetrafluoroethyleneProtein Interaction MappingProteinsViral ProteinsViroporin ProteinsBiopolymersCarbon IsotopesHydrogenNitrogen Isotopesp7 protein, Hepatitis C virusPolytetrafluoroethyleneProteinsViral ProteinsViroporin Proteins

Identifiers

PMID19559634
PMCPMC2741312
OpenAlexW1995855668

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.