Assessing the performance of handheld glucose testing for critical care.
Article in Diabetes technology & therapeutics, 2008. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
7 citing papers in PubMed, 36 citations in OpenAlex.
- Using geographic rescue time contours, point-of-care strategies, and spatial care paths to prepare island communities for global warming, rising oceans, and weather disasters.International journal of health geographics · 2023Article
- Utility of point-of-care vs reference laboratory testing for the evaluation of glucose levels.Diabetic medicine : a journal of the British Diabetic Association · 2019Article
- Article
- Computing the surveillance error grid analysis: procedure and examples.Journal of diabetes science and technology · 2014Article
- The StatStrip glucose monitor is suitable for use during hyperinsulinemic euglycemic clamps in a pediatric population.Diabetes technology & therapeutics · 2014Article
- Analytical and clinical performance of blood glucose monitors.Journal of diabetes science and technology · 2010Review
- Point-of-care testing for disasters: needs assessment, strategic planning, and future design.Clinics in laboratory medicine · 2009Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors at 2 institutions in 1 country.
Funding
Abstract
backgroundWe assessed the performance of a point-of-care (POC) glucose meter system (GMS) with multitasking test strip by using the locally-smoothed (LS) median absolute difference (MAD) curve method in conjunction with a modified Bland-Altman difference plot and superimposed International Organization for Standardization (ISO) 15197 tolerance bands. We analyzed performance for tight glycemic control (TGC).
methodsA modified glucose oxidase enzyme with a multilayer-gold, multielectrode, four-well test strip (StatStriptrade mark, NOVA Biomedical, Waltham, MA) was used. There was no test strip calibration code. Pragmatic comparison was done of GMS results versus paired plasma glucose measurements from chemistry analyzers in clinical laboratories. Venous samples (n = 1,703) were analyzed at 35 hospitals that used 20 types of chemistry analyzers. Erroneous results were identified using the Bland-Altman plot and ISO 15197 criteria. Discrepant values were analyzed for the TGC interval of 80-110 mg/dL.
resultsThe GMS met ISO 15197 guidelines; 98.6% (410 of 416) of observations were within tolerance for glucose <75 mg/dL, and for > or =75 mg/dL, 100% were within tolerance. Paired differences (handheld minus reference) averaged -2.2 (SD 9.8) mg/dL; the median was -1 (range, -96 to 45) mg/dL. LS MAD curve analysis revealed satisfactory performance below 186 mg/dL; above 186 mg/dL, the recommended error tolerance limit (5 mg/dL) was not met. No discrepant values appeared. All points fell in Clarke Error Grid zone A. Linear regression showed y = 1.018x - 0.716 mg/dL, and r2 = 0.995.
conclusionsLS MAD curves draw on human ability to discriminate performance visually. LS MAD curve and ISO 15197 performance were acceptable for TGC. POC and reference glucose calibration should be harmonized and standardized.