ArticleBritish journal of clinical pharmacology1999
A comparative analysis of the particle size output of beclomethasone diproprionate, salmeterol xinafoate and fluticasone propionate metered dose inhalers used with the Babyhaler, Volumatic and Aerochamber spacer devices.
Article in British journal of clinical pharmacology, 1999. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed, 39 citations in OpenAlex.
- Comparison of adverse events associated with different spacers used with non-extrafine beclometasone dipropionate for asthma.NPJ primary care respiratory medicine · 2019Article
- A review of the in vitro and in vivo valved holding chamber (VHC) literature with a focus on the AeroChamber Plus Flow-Vu Anti-static VHC.Therapeutic advances in respiratory diseaseReview
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
2 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
aimsTo determine in vitro the effect of delay, inspiratory flow, and spacer washing on the drug output of metered dose inhalers (MDIs) used with different spacer devices.
methodsThe amount of drug in particles <5 microm diameter from MDI+spacer, sampling after a delay of up to 20 s, was measured using a Multistage Liquid Impinger. Drug output was also measured at different flow rates, and after washing the Babyhaler in household detergent.
resultsMore fluticasone in small particles was recovered from the Babyhaler than the Volumatic or the Aerochamber spacers, and more beclomethasone and salmeterol was recovered from the Babyhaler and Volumatic spacers than from the Aerochamber. Washing the Babyhaler reduced the recovery of salmeterol, and did not alter the recovery of the other drugs tested.
conclusionsSpacer devices need to be fully evaluated for each drug prescribed for them.